Function

 

3.

The portable JCP-300 and JCP-310  sample gas condi-
tioning systems are built in a robust carrying and mobile 
unit case made of reinforced plastic. All parts needed for 
operation or maintenance are placed on the front panel. 
Gas connections are generally on the front panel. 
 
 

Gas heat exchanger 

3.1.

The JHEX-4 heat exchanger consists of a cylindrical 
body. The gas flow is designed to bring the wet gas as 
often as possible in contact with the cooled surfaces. 
The manner in which the gas circulates causes laminar 
as well as turbulent flow that enhances the separation of 
the condensate. The outside of the heat exchanger is 
thermo-insulated. This prevents the condensation of 
humid air on the outside of the heat exchanger. 
 

 

Cooling 

3.2.

For cooling of the gas heat exchanger Peltier elements 
are used. This element is mounted in the upper part of 

 

Funktion

 

3.

Die tragbaren Messgasaufbereitungen JCP-300 und 
JCP-310 sind in ein robustes trag- und fahrbares Kunst-
stoffgehäuse eingebaut. Sämtliche Bedienungs- oder 
Wartungselemente sind auf der Gerätefront angeordnet. 
Die Gasanschlüsse befinden sich bei den Standard-Mo-
dellen in der Geräteseitenwand. 
 

Gaswärmetauscher 

3.1.

Der JHEX-4 Wärmetauscher besteht aus einem zylindri-
schen Hohlkörper. Die Gaswege im Wärmetauscher 
sind so gestaltet, dass das feuchte Gas in Kontakt mit 
der gekühlten Oberfläche tritt. Aufgrund der Gasführung 
bilden sich sowohl laminare wie auch turbulente Strö-
mungen, welche die Kondensatabscheidung begünsti-
gen. Der Wärmetauscher ist thermisch isoliert, dadurch 
wird eine Kondensation von Luftfeuchtigkeit an der Wär-
metauscheraußenseite verhindert. 
 

Kühlung 

3.2.

Für die Kühlung des Gaswärmetauschers kommen Pel-
tier-Elemente zum Einsatz. Ein durch den Peltiereffekt 

Manual JCP-300 Series

BA_DE_JCP-300_v1.0

          

–––––––––––––––––––––––––– [ 5 ] ––––––––––––––––––––––––––

JCP-310 high performance 

including JSR-25 condensate pump

Heat exchanger

JHEX-4 heat exchanger aluminum coated

1

JHEX-4 heat exchanger PVDF

2

JHEX-4 heat exchanger Duran glass

3

JHEX-4 heat exchanger stainless steel

4

Filtration

without

0

fine dust filter JF-1

1

Sample gas pump

without

0

sample gas pump JSP

1

Flow meter with needle valve

without

0

flow meter with needle valve

1

Condensate detector

without

0

condensate detector

1

Temperature display for 
heat exchanger

without

0

temperature display for heat exchanger

1

Acid dosing pump

without

0

acid dosing pump

1

Temperature controller 
for heated sample line

without

0

temperature controller for heated sample line Pt100

1

temperature controller for heated sample line TC type K

2

Power supply

230 VAC 50/60 Hz

A

115 VAC 50/60 Hz

B

Special version

XY

Order code

                                                                   JCP-310.

Содержание JCP-300

Страница 1: ...JCT ANALYSENTECHNIK GMBH BEDIENUNGSANLEITUNG OPERATING MANUAL 3 20 JCP 300...

Страница 2: ...drain 12 6 5 Leakage test 14 6 6 Electrical connections 14 6 7 Connector plug terminal strip 14 7 Start up 15 8 Operation 16 9 Temperature controller heated sample line 16 9 1 General 16 9 2 Display o...

Страница 3: ...d Die JCT Analysentechnik GmbH haftet nicht bei eigen m chtigen nderungen des Ger tes oder f r deren un sachgem e Handhabung oder Verwendung Wenn anzunehmen ist dass ein gefahrloser Betrieb des Ger te...

Страница 4: ...ystemen bestimmt Beachten Sie die Angaben in den technischen Spezifikationen hinsichtlich Umge bungs und Versorgungsbedingungen sowie zul ssige Druck und Temperaturgrenzen Dieses Ger t darf nicht in e...

Страница 5: ...aus einem zylindri schen Hohlk rper Die Gaswege im W rmetauscher sind so gestaltet dass das feuchte Gas in Kontakt mit der gek hlten Oberfl che tritt Aufgrund der Gasf hrung bilden sich sowohl lamina...

Страница 6: ...s pump 3 6 The sample gas pump forwards the sample gas through erzeugter W rmetransport entzieht ber den Gasw r metauscher dem zugef hrten Messgas die Energie und f hrt diese durch die K hlrippen ab V...

Страница 7: ...ng can be equipped with a temperature controller Manual JCP 300 Series aufbereitung zum nachgeschalteten Analysator Sie kann durch den Kippschalter in Betrieb gesetzt werden und ist durch die optional...

Страница 8: ...with trolley function polypropylene yellow Gas wetted parts depending on heat exchanger aluminium coated PVDF SS316Ti FFKM Duran glass Dead volume per gas path 67 ml Connection sample gas inlet SS316...

Страница 9: ...K1233009A Rollentr ger Kondensatpumpe K1233011 Laufband 17 00002 Filterelement Glasfaser 5 Stk 17 00003 Filterelement PTFE 3 Stk 17 90001 O Ring Dichtung Filtergeh use VPE 3 Stk 12 90431 Verschlei te...

Страница 10: ...le gas inlet temperature 140 C Ersatzteile K1004600 Ventilator 230 VAC K1004605 Ventilator 115 VAC K1204323 Peltier Regelelektronik JPCU 1 K1204360 Kondensatw chter Elektronik KW 2 17 04000 Kondensats...

Страница 11: ...sensor KW 1 Option 5 Condensate sensor KW 1 option Messgaspumpe Option 6 Sample gas pump option Feinstaubfilter Option 7 Fine dust filter option Durchflussmesser mit Nadelventil 8 Flow meter with need...

Страница 12: ...ist das Ger t auf allf llige Trans portsch den zu untersuchen Wurde ein Schaden fest gestellt sind unverz glich die verantwortliche Spedition und der H ndler zu benachrichtigen Es ist zu berpr fen ob...

Страница 13: ...e line Filterelement 3 Filter element Kondensatsensor KW 1 4 Condensate sensor KW 1 Durchflussmesser mit Nadelventil 5 Flow meter with needle valve Temperaturregler f r beheizte Messgasleitung 6 Tempe...

Страница 14: ...g 6 5 VORSICHT Nach dem Anschluss ist die Dichtheit aller Schlauchver bindungen zu pr fen Elektrischer Anschluss 6 6 rtliche Netzspannung Netzfrequenz und Leis tungsaufnahme mit den Angaben am Typensc...

Страница 15: ...schl sse VORSICHT Betriebsspannung und max zul ssige Stromaufnahme von 8A beachten Heizleitung nur spannungslos an schlie en VORSICHT Dieses Ger t wird mit Netzspannung betrieben Beim Betrieb stehen z...

Страница 16: ...Manual JCP 300 Series 7 Energieversorgung herstellen Das Ger t beginnt zu k hlen Der Ventilator muss drehen und die Peltier Status LED leuchtet f r ca 15 min rot und die optionale Temperaturanzeige z...

Страница 17: ...den am Ger t f hren BA_DE_JCP 300_v1 0 17 Betriebszustand Status Status LED Condensate Gr n 1 green 1 Status LED Condensate Rot 1 red 1 Status LED Condensate blinkt 1 blinking 1 Status LED Peltier Gr...

Страница 18: ...arameter Wert werden angezeigt Dieser Modus erlaubt eine Einstellung von SV und des Alarmwertes 2 Dr cken der Block Taste um Setup Modus zu akti vieren Der Setup Modus erm glicht die Einstellung der P...

Страница 19: ...tions Remove probe from process flange Manual JCP 300 Series 5 Einstellung des Temperatur Sensors Zwei Varianten von Temperatursensoren werden ge meinsam mit dem JPXE4 verwendet 2x dr cken um Setup Ch...

Страница 20: ...iliar with the dangers involved Manual JCP 300 Series Anschluss des Kondensatablaufs trennen Fachgerechte Aufbewahrung bzw Entsorgung Recycling Das Ger t enth lt Bauteile die wiederverwertet werden k...

Страница 21: ...geltenden Umwelt Sicherheits und tech nischen Richtlinien zu erfolgen Reinigen der K hllamellen 12 1 Zum Reinigen der K hllamellen den r ckw rtigen De ckel abnehmen und Lamellen mit Druckluft freibla...

Страница 22: ...clockwise ro tation Schritt 2 Step 2 Die beiden Kondensat pumpen schl uche von den Fittings abziehen Pull off both condensate pump tubes from fittings Schritt 3 Step 3 Verschlussriegel im Uhrzeigersin...

Страница 23: ...bands am Pumpenkopf montie ren und beide Enden einschnappen Mount the tubing cover on the pump head place both end parts in the rail until they snap in Schritt 6 Step 6 Verschlussriegel gegen den Uhrz...

Страница 24: ...e two screws for pump head fixing and pull pump head with pulley holder off Schritt 2 Step 2 Pumpkopf mit neuem Rollentr ger auf Achse aufschieben Wellenschulter muss nach vorne zeigen Achten Sie dara...

Страница 25: ...t gefroren Umgebungstemperatur 1 C ggf Betriebsraum beheizen Kondensat alarm Ursache beseitigen siehe feuchtes Messgas Kondensatsensor durch L sen der ber wurfmutter nach unten abziehen reinigen und t...

Страница 26: ...DIN 34 Schutzvermerk nach DIN 34 beachten The technical data and design disclosed Analysentechnik GmbH and or contain propriertary rights of others and are not to be use sap 04 03 2020 Benennung P340...

Страница 27: ...tokoll 15 BA_DE_JCP 300_v1 0 27 T tigkeit Activity durchgef hrt von performed by Datum Date Unterschrift Signature K ltemittelverfl ssiger reinigen Condenser cleaned Pumpenschl uche der Kondensat pump...

Страница 28: ...AATWWXXX IBAN AT25 2011 1829 4301 5600 UNICREDIT BANK AUSTRIA AG BIC BKAUATWW IBAN AT89 1100 0039 9372 2200 UID Nr VAT No ATU20559900 Kreisgericht Wr Neustadt FN 119682 d Gesch ftsf hrer CEO G nter Zi...

Отзывы: