PROCEDURE FOR SYSTEM ADJUSTMENT 

 

 
Should users wish to verify the output temperatures of JBC Advanced Series equipment we suggest using our 

TI-A Thermometer

 and the “

Temp Adjust

” mode by using the internal menu settings of the control units. 

 
 

1)  Turn on  the station to be adjusted and 

choose the working temperature in the 
station. 

 

2)  Take the soldering iron and, when the 

temperature is steady, apply fresh solder to 
the tip of the soldering iron. 

 

3)  Clean the sensor of the thermometer from 

previous flux residues if necessary. 

 

4)  Place  the tip on the sensor of the 

thermometer  by  forming an angle of 30ºC 
as shown in the picture. 

 

5)  Check the solder on the tip is doing right 

thermal contact to the sensor. If necessary, 
add new fresh solder. 

 

6)  Wait until the temperature is steady. 

 

7)  To correct  the possible error between the temperature shown by the display of the station, and the 

temperature measured in the thermometer, access the “temp adjust” option, inside the “tool settings”, 
in the main menu of the station. 

 

8)  Here you can modify the temperature of the tip by using the buttons of the station. For example, if you 

see 350 degrees in the display, and 360 in the thermometer, you have to set -10 degrees in the “temp 
adjust” option in order to adjust the tip temperature.  

 

9)  In the opposite case, if you read 350 degrees in the display, and 340 in the thermometer, you have to 

set +10 degrees in the “temp adjust” option. 

 

10) Save parameters and exit the menu. 

 

11) Repeat from step 1 to 5 to verify. 