1
USER GUIDE
OptiCell 4 PRO
air mattress system
V2
Страница 1: ...1 USER GUIDE OptiCell 4 PRO air mattress system V2 ...
Страница 2: ... QUALITY Our quality standards and customer focus goals are realised by responding to the needs and expectations of our customers We believe in building long term relationships through collaboration and trust Continuous improvement in our products and services is the basis of our ability to satisfy both current and future customers Our approach includes the development of evaluation methods and st...
Страница 3: ...attress system 5 Preparation 5 The OptiTex cover 6 Intended use 6 Fire protection 6 Features OptiCell 7 Use of product 8 Operation 9 Air hose connection 10 Hand Check test 11 Inspection and troubleshooting 12 Cleaning 13 Pump overview OC30 2 0 14 Specifications 15 Recycling 16 Other information 17 Accessories 18 Symbols 19 ...
Страница 4: ... of the intended usage environment and then use the right product Please contact us if you have any questions In order to avoid damage the mattress should not come into contact with heat sources or fire such as burning cigarettes heat lamps or the like Make sure that you use the correct mattress size for the bed and the appropriate model for the patient Connect the product to the nearest situated ...
Страница 5: ...RATION This user guide contains information for general use maintenance and improved safety It is important to read this guide carefully before using the product Care personnel must be given access to this user guide and also be informed of the risks that may arise when using electrical equipment Training on the product can be provided on request Before using the product for the first time or when...
Страница 6: ...nt Fluid protection flap and waterproof zip on four sides All seams are welded Anti slip underside which grips to bed base in addition to four fastening straps Four transportation handles to enable easy moving INTENDED USE The OptiCell 4 PRO V2 has a height of 20cm and is an alternating pressure air mattress which is placed directly onto a bed base It aids the prevention and treatment of pressure ...
Страница 7: ...he OptiCell 4 PRO V2 air mattress is connected to the pump by three air hoses The red air hose positioned in the middle is used to fill the head cells situated under the patient s head It also fills the lower chambers of the standard cells and the ventilation cells This ensures that there is always air in the mattress no matter how the pump is set or if the pump should stop working for example due...
Страница 8: ...p in mind that a tall patient can weigh the same as a short patient but may need a different pressure This must always be carefully observed for each patient The biggest risk for the patient is to bottom out on the mattress The basic rule is that the part of the body to be unburdened must not be subject to pressure If the patient bottoms out the objective of the mattress is lost See page 11 for in...
Страница 9: ...nt to be disturbed by an alternating pressure Press the Care Static button position E on page 14 to select either Care mode or Static mode Care also called Auto firm mode provides maximum pressure to facilitate nursing care Static adjusts the air pressure after setting the comfort mode The correct therapeutic effect is achieved with alternating air pressure in Dynamic mode Static air pressure cons...
Страница 10: ...e connector if the air hose is disconnected from the pump To minimise the risk of air leakage during hose connection check That the two hooks are not damaged and their clips are intact That the black o rings on the connector s three air hoses are not cracked or missing That no debris has entered the connector causing a bad connection or blocking the air hoses The air hose connector is fitted by a ...
Страница 11: ... that it is easier for the staff but it may give an incorrect test result Fold away the sheets and unzip the mattress cover to access the cells Put one hand between two cells with the patient lying on the mattress See illustration Depending on the size of the hand it may sometimes be sufficient to use 2 3 finger height Ensure that when the hand is in position it fits between the bed base and the p...
Страница 12: ...ir hoses for signs of damage or abnormal wear Check the air hose connector see page 10 INSPECTION OFTHE OPTITEX COVER Regularly check the cover for visible signs of damage or wear such as tears cracks holes stains or discolouration Check all sides including the underside TROUBLESHOOTING The pump does not work or does not start Make sure the electrical plug is properly inserted then turn on the pum...
Страница 13: ...t a minimum of 70 C for at least 10 minutes with unscented detergent according to standard SS EN 14065 2016 Textiles Laundry Processed Textiles Biocontamination Control System The OptiTex cover can be machine washed at a maximum of 95 C CLEANING THE CELLS AND AIR HOSES The cells can be washed separately or in groups using a damp cloth and mild detergent The air hoses can also be wiped with a wet c...
Страница 14: ...the panel press and hold the button for 5 seconds Mute alarm If the alarm sounds press this button to mute and reset The adjacent lamp will light up if the air pressure in the mattress is too low Check if necessary for any air leakage see page 12 The adjacent lamp marked will light up in the event of a power failure Cycle time for alternating pressure Use this button for setting the preferred cycl...
Страница 15: ...S Pump OC30 2 0 Specifications Size 28 0 x 11 0 x 20 5cm Weight 2 6 kg Pressure range 20 70mm Hg Cycle time 10 15 20 or 30 minutes System Specifications Classification Class II IPXO No AP APG Type BF Applied part Mattress Power rating 230V AC 50 Hz 1A 12W Fuse rating 250V T 1 A Operating environment Temperature Use 10 35 C Storage 15 50 C Transport 15 70 C Humidity Use 20 80 without condensation S...
Страница 16: ...sposed of at a recycling station and sorted as Combustible waste For more information contact Järven HealthCare Järven Health Care takes environmental responsibility seriously We strive to always use materials and products with the least possible impact on the environment We are constantly working to develop and improve production techniques Järven Health Care is affiliated with the Swedish FTI an...
Страница 17: ...d service life of our OptiCell systems is 5 years RETURNS AND COMPLAINTS Any complaints regarding the product must be notified in writing to Järven Health Care The return of products must be done in accordance with our instructions Returning a product before a complaint case is processed will result in the case being considered closed and customer compensation will not be paid DAMAGE UPON DELIVERY...
Страница 18: ...lon PU cell ventilating OC 2 NX 2 PRO and Gentle 90 cm 12 960402 004 Nylon PU cell ventilating OC 2 NX 2 PRO and Gentle 105 cm 12 960402 005 Nylon PU cell ventilating OC 2 NX 2 PRO and Gentle 120 cm 12 970103 008 OptiTex hygienic cover 4 NX 200x 90x20 cm OC 2 PRO 3 PRO 4 NX 4 PRO and Gentle 12 970103 009 OptiTex hygienic cover 4 NX 200x105x20 cm OC 2 PRO 3 PRO 4 NX 4 PRO and Gentle 12 970103 010 O...
Страница 19: ...ich pressure ulcer category the mattress is suitable for Indicates the product s article number Indicates the product s batch number Indicates serial number Date of manufacture Manufacturer The product is a medical device Disinfection should be undertaken with alcoholbased or similar surfactantbased detergents No chlorine containing preparations may be used Tumble dry at low temperature Plastic mu...
Страница 20: ...t to update this guide if necessary and or on finding any spelling or factual errors Additional copies can be ordered from Järven Healthcare Copying for internal use is allowed OptiCell 4 PRO V2 rev 22 01 00 OptiCell AB Bromsvägen 3 SE 891 60 Örnsköldsvik SWEDEN Tel 46 660 22 28 50 Fax 46 660 824 10 Email healthcare jarven se ...