Jaquet DualTach T501 Скачать руководство пользователя страница 1

Instruction Manual T501 

 

JAQUET AG 

 

 
Version 2.00E 

 

1/59 

 

 

 

Frequency Measurement and Switching Instrument 

DualTach T501

 

Instruction Manual 

Version 2.00E 

 

T501 : 2 Channel Tachometer with 4 Relays and 2 Analog Outputs 0/4-20mA

 

 

T501.50: Part No.: 384Z-05600 (AC Version) 

 

T501.10: Part No.: 384Z-05601 (DC Version) 

 

 

Содержание DualTach T501

Страница 1: ...E 1 59 Frequency Measurement and Switching Instrument DualTach T501 Instruction Manual Version 2 00E T501 2 Channel Tachometer with 4 Relays and 2 Analog Outputs 0 4 20mA T501 50 Part No 384Z 05600 AC...

Страница 2: ...ion 12 3 5 1 Ethernet 12 3 6 Environment 12 3 6 1 Climatic Conditions 12 3 6 2 Electromagnetic immunity 12 3 6 3 Other Standards 13 4 PRINCIPLE OF OPERATION 14 4 1 General 14 4 2 Machine factor 16 4 2...

Страница 3: ...Sensor alarm 32 7 7 5 Machine factor 32 7 7 6 Dynamic Error 33 7 7 7 Math function 34 7 7 8 System Limit 35 7 7 9 Relay outputs 36 7 7 10 Open Collector Outputs 37 7 7 11 Analog output 38 7 7 12 Copy...

Страница 4: ...Power failure 47 8 6 Behaviour during Configuration 47 9 MECHANICAL CONSTRUCTION HOUSING 48 9 1 Tachometer 48 9 2 Mounting 49 10 ACCESSORIES 50 11 MAINTENANCE REPAIR 50 12 SOFTWARE VERSION 50 12 1 Co...

Страница 5: ...rs may only be opened for repair by trained personnel These Instruments correspond to protection class I The PE terminal must therefore be earthed The instructions in this manual must be strictly adhe...

Страница 6: ...eed sensors 2 Analog outputs 2 Binary input 4 Relays 2 Open Collector Outputs 4 Parameter sets each having 6 System limits with 4 Limit values Sensor monitoring System monitoring The Tachometers are c...

Страница 7: ...t 2 Relay outputs Open Collector output 1 Open Collector output 2 Ethernet interface Power consumption P Min W P Max W DC Version 18 VDC 3 3 6 6 24 VDC 3 4 6 7 36 VDC 3 5 6 8 AC Version 110 VAC 5 0 11...

Страница 8: ...Min Pulse width 5 s NO DC decoupling Input signal for adaptive trigger min 57mVpp 10 20mVRMS Max Voltage 80VRMS 226Vpp DC decoupling The Trigger level automatically adapts itself to the input signal...

Страница 9: ...1 1 10 100 1000 10000 100000 Frequency Hz Sensor Signal V pp O K NOT O K Sensor supply 14V 0 5V max 35mA Short circuit proof If the current limit has been activated then the circuit needs to be disco...

Страница 10: ...mpedance is dynamically measured Z 60 k OK Z 125 k NOT OK 3 3 Binary Inputs Number 2 Active Binary Inputs and 1 push button EXEC logic OR with Binary input B1 Level Logic 0 0V 5V Logic 1 15V 36V Isola...

Страница 11: ...7uA Measuring error Max 0 15 f m v 2 LSB 25 C 50 C Max 0 20 f m v 2 LSB 40 C 70 C Accuracy class 0 1 of the end value Damping Hardware time constant T 4 1 ms 10 90 8 6 ms Hardware Actual time Hardwar...

Страница 12: ...put signal 3 5 Data communication 3 5 1 Ethernet Number 1 Physical Layer Ethernet 10Base T IEEE 802 3i Max cable length 100 m Transmission rate 10 MBit s Connection Front panel 8P8C RJ45 Usage Configu...

Страница 13: ...Instruction Manual T501 JAQUET AG Version 2 00E 13 59 3 6 3 Other Standards Communaut Europ enne...

Страница 14: ...be individually defined Parameter sets may be selected via binary inputs The 4 relays and 2 open collector outputs may be assigned to any System Limit and will react accordingly The 2 open collectors...

Страница 15: ...Instruction Manual T501 JAQUET AG Version 2 00E 15 59...

Страница 16: ...hysical parameters In principle any parameter may be processed as long as it can be translated into a frequency proportional value The formulae above may still be applied whereby the required paramete...

Страница 17: ...correspond to protection class I The PE terminal must therefore be earthed Before switching on check that the supply is within the permissible range The Sensor cable screens must be connected terminal...

Страница 18: ...0E 18 59 6 Connections 6 1 Front view 6 1 1 Front view T501 The Ethernet Interface status LED along with the EXEC push buttong are located at the front For communications please see Paragraph 7 Config...

Страница 19: ...sor connections S1 S2 S1 Sensor input S1 Sh1 Screen S1 V1 Sensor Supply S1 V1 0V S1 S2 Sensor input S2 Sh2 Screen S2 V2 Sensor Supply S2 V2 0V S2 Analog outputs I1 positive I1 I1 negative I1 I2 positi...

Страница 20: ...DSD electromagnetic 2 wire Sensors DSE Green Line E Sensors with amplifier and fixed Trigger level at 1 25 3 0 Volt DC DSF DSD Greenline F D Y or Sensor types or configuration recommendation of the S...

Страница 21: ...er Oc1 negative Open Collector 1 Collector Oc2 positive Open Collector 2 Emitter Oc2 negative Open Collector 2 Collector Relay Outputs R1 R4 NC Normally closed NO Normally open Com Common Oc2 R4 com R...

Страница 22: ...ocedures apply and the configuration file can be communicated between computer and T501 To run the Software you must have Java Runtime Environment JRE 1 5 or higher 7 2 PC Settings The PC Ethernet car...

Страница 23: ...ependent upon your Internet Explorer configuration you will be asked whether you want to Open Run or Save Save as the programme 7 3 1 Direct execution Select Open and the configuration will be read fr...

Страница 24: ...cess and control over all Parameters Process as well as Configuration The factory settings are shown below Bold 7 4 2 Log in Here you can Log in on three different levels The standart passwords for Co...

Страница 25: ...Instruction Manual T501 JAQUET AG Version 2 00E 25 59 7 4 4 Logging in and out To change user level log out via File and click on Log out then log in again using the appropriate password...

Страница 26: ...ved the live parameters then they would be lost 7 5 2 resetting to factory default To reset parameters to factory settings go to Settings Back to factory setup then click on OK NB if you have not save...

Страница 27: ...ving To save actual parameters go to File Save as and choose the path and file name you want to use Please note that the file name ends in the format T501 7 5 5 printout To print actual parameters go...

Страница 28: ...h interval defined under Settings Refresh Interval Second Second 1 Second 2 Seconds 5 Seconds 10 Seconds To end the measured data display go to Online Stop read measure data 7 6 2 Reading configuratio...

Страница 29: ...er can take 10s If no connection can be established this will be aborted after 3 attempts NB The actual file in the tachometer will be overwritten 7 6 4 Compare data If you would like to compare a PC...

Страница 30: ...pply To use static sensor monitoring the Min Max current consumption must be defined Not powered The sensor is not powered by the T501 and no static monitoring is possible Internal pull up resistor De...

Страница 31: ...t 15V 33V Active Low Logic 1 binary input 0V 5V Binary input for System Limit input Binary input is a constituent part of System Limits Parameter set selection Binary inputs select the parameter set P...

Страница 32: ...sensors are present see Paragraph 7 7 6 Dynamic Error 7 7 5 Machine factor Go to Configuration System The Machine factor establishes the relationship between sensor frequency and associated machine s...

Страница 33: ...n System f f S S Differenz 2 1 fS1 Sensor 1 frequency fS2 Sensor 2 frequency This function is only available used when both sensor inputs are used This function has influence only on Sensor alaram Dev...

Страница 34: ...S2 S1 100 x S1 S2 S2 Percentage difference 100 x S2 S1 S1 Variance S1 Variance Variance S2 Acceleration S1 Speed value acceleration Acceleration S2 S1 The speed values are fed through e g for the Wind...

Страница 35: ...r 2 Limit low 0 01 400 0 999 999 Limit high 0 01 500 0 999 999 Math Function Limit low 999 999 50 0 999 999 Limit high 999 999 100 0 999 999 Inversion Sensor 1 Over speed Under speed Sensor 2 Over spe...

Страница 36: ...and the Fail safe Not fail safe mode Possible relay assignment Latch Function Latched not latched Safety function Fail safe Not fail safe System limit 1 Sensor Error System limit 2 Static S1 System li...

Страница 37: ...mic error System Limit 3 Always ON System Limit 4 Always OFF System Limit 5 Frequency S1 System Limit 6 Frequency S2 Sensor Error Frequency x2 Static S1 Frequency x4 Latch Function Latched not latched...

Страница 38: ...999 0 0 999 999 End value 999 999 2 000 0 999 999 Time constant 0 9 9 in 0 1 Second increments Assignment selection The selection box is used to choose Analog Output assignment Current range 0 20mA or...

Страница 39: ...Inclusion of input in the logical combination Inclusion of Sensor alarm in the logical combination Relay Open Collector assignment Relay Open Collector latch mode Relay Open Collector fail safe mode A...

Страница 40: ...sword or Change Process user password Changing the password involves entering the old password and the new password twice To save the password click on OK The password is now stored in the configurati...

Страница 41: ...idual Process name may be entered into the text field to the right of Prozess ID max 16 characters Only alphanumeric characters Type Tachometer type number Java version Java Version stored in the Tach...

Страница 42: ...tput Up until the first measurement interval is completed all relays are de energised Thereafter they assume the defined condition If no input frequency is present then after measurement interval x 2...

Страница 43: ...f measurement intervals without input signal This continues until a second negative edge arrives Input signal Input period Measurement End of the measurement interval Time The calculation and adjustme...

Страница 44: ...sets A B C and D The binary inputs must be used to change parameter sets To configure binary inputs see paragraph 0 Binary input configur Binary input configuration Status BIN 1 Status BIN 2 Selected...

Страница 45: ...To set a Window function on a speed input please follow the steps below Example for Sensor 1 1 Set the Math function to S1 2 Set Math function and Sensor 1 hysteresis 3 The Sensor hysteresis gets the...

Страница 46: ...of 35kHz Frequency S1 Frequency S2 Frequency x2 Frequency x4 90 10us 8 3 6 Relay and Open Collector Latch Function Relays and Open Collectors can be assigned a latch function If a signal arrives from...

Страница 47: ...ent of a sensor fault is a function of the software configuration If sensor monitoring is selected a fault causes the LED to go off and the analog output both go to error status The Relay and open col...

Страница 48: ...Instruction Manual T501 JAQUET AG Version 2 00E 48 59 9 Mechanical Construction Housing 9 1 Tachometer...

Страница 49: ...Instruction Manual T501 JAQUET AG Version 2 00E 49 59 9 2 Mounting Wall mounting Top hat rail mounting as delivered...

Страница 50: ...leaning may be carried out using spirit pure alcohol or soap only 12 Software Version History and changes of the configuration software and firmware 12 1Configurations software Software version Change...

Страница 51: ...Instruction Manual T501 JAQUET AG Version 2 00E 51 59 Appendix A Declaration of Conformity...

Страница 52: ...le Select Deactivate Re open Menu File Select Activate Yellow LAN connector LED lights but no connection to browser Cause Several instruments are connected to the same LAN as the tachometer Solution T...

Страница 53: ...t protocol TCP IP and select the options IP Address automatically recognize DNS Server address automatically recognize Then select Advanced Insure that apart from the entries shown no other entries ar...

Страница 54: ...t the tag WINS and insure that no entries are present NetBIOS Setting is Standard Confirm with OK Change to tag Alternative Configuration Insure that Automatically allocated private IP Address is sele...

Страница 55: ...phics Cause Your browser and computer interpret the address incorrectly Solution Extend the address with a forward slash http 192 168 1 127 software Problems when using Microsoft Internet Explorer Cau...

Страница 56: ...ange setting to Medium Click on Reset Click on OK Click OK again Start IE again and proceed as described in the operating instructions The first page is shown correctly but after selecting Download no...

Страница 57: ...ating instructions Solution Should you not be able to change these settings we recommend you try the browser Firefox freeware Problems using Internet Browser Mozilla Cause Ping works but an error mess...

Страница 58: ...nual T501 JAQUET AG Version 2 00E 58 59 Select Direct Connection to Internet Confirm with OK How do I see which Java Version is installed on my PC Solution Win XP Go to Control Panel And select Java t...

Отзывы: