background image

Safety information

Please read the operating instructions enclosed 
diligently, as well as all other publications that 
must be drawn upon for working with this 
measurement case and the accessories.

Observe all safety instructions and warnings. 
Failure to comply with the instructions can result 
in personal injuries and/or damage to the product.

The measurement case must only be operated by 
specialised personnel. Specialised personnel are 
persons, that based on their respective training 
and experience, are qualified to recognise risks 
and prevent potential dangers that can be 
caused by the operation or maintenance of the 
measurement case.

Check the measurement case, the measurement 
wires and accessory parts for any signs of 
damage before use. Check whether there are 
cracks in the plastic of any parts missing. Pay 
particular attention to the insulation of the wiring 
and connectors. Defective parts must be replaced 
immediately. Do not use if damaged! 

Comply with all of the locally applicable and 
national safety regulations. 
Use personal protective equipment (e.g. approved 
rubber gloves, face protection and protective 
clothing) in order to avoid electric shocks and 
burning injuries.

If the measurement case is not operated in 
accordance with the operation manual, protection 
is no longer ensured and hazards can be 
presented by the device!

Do not work with the device if the front plate 
has been removed. Otherwise, you may come 
into contact with hazardous voltages! It is not 
permitted to use the device with the front plate 
removed!

The UMG 96RM-E is not suitable for 
measuring DC voltages.

Measured voltages and measured 
currents must derive from the same 
network. Voltages outside the rated 
range could destroy the measurement 
case.

Incorrect network settings can cause 
faults in the network.

m

- 3 -

Do not use the measurement case in 
wet or moist environments or in the 
vicinity of explosive gases or vapours.

Use the measurement case only with 
the measurement wires and adapter 
supplier.

Exercise particular caution when 
connecting and removing the flexible 
current transformers and the gripping 
probes for the measured voltage. 
Ensure that the circuit to be measured 
is de-energised.

Keep your fingers behind the finger 
protectors on the gripping measurement 
probes.

Voltages and currents that are outside 
the permissible measurement range
can destroy the device.

The input voltage shall not be 
higher than the rated voltage of the 
measurement device.

Disconnect the input signals before 
cleaning the measurement case.

c

c

Attention! Fatal hazard due to electrical energy: 

High voltage spikes that are dangerous to touch can occur on current transformers that 
are driven with open-circuit secondary windings! For this reason, never operate current 
transformers with open-circuit secondary windings! 

First connect the current transformer lines to the corresponding measurement case inputs!

Содержание MRG 96RM-E Flex RCM

Страница 1: ...bH Vor dem Polst ck 6 D 35633 Lahnau Support tel 49 6441 9642 22 Fax 49 6441 9642 30 E mail info janitza com Internet http www janitza com www janitza com Mobile universal measurement device MRG 96RM...

Страница 2: ...basic device 4 Connecting the measurement case 4 Reading out recordings 5 MRG 96RM E Flex RCM views 5 Measurement transformer settings 6 Example connection voltage measurement 7 Example connection cu...

Страница 3: ...Do not work with the device if the front plate has been removed Otherwise you may come into contact with hazardous voltages It is not permitted to use the device with the front plate removed The UMG 9...

Страница 4: ...ered by this operation manual for the UMG 96RM E please contact the support department of Janitza electronics GmbH Tel no 49 6441 9642 22 This device must only be operated and repaired by specialised...

Страница 5: ...MRG 96RM E Flex RCM views 5 Power Analyser UMG 96RM E Measurement transducer Current transformer inputs I L1 I L2 I L3 I N Supply voltage AUX 24V DC 1A I L1 Current transformer inputs UMG 96 RM L1 L2...

Страница 6: ...A 6 630 A 5 1000 A 4 1500 A 3 2000 A 2 4000 A 1 Setting Rogowski coil 300 mm Flex CT 1A 300 mm Set the current measurement range with the DIP switch 2 in accordance with the table Potentiometer 1 DIP...

Страница 7: ...fuse Max 20mm Locking sleeve with bayonet catch for accommodating a fuse Example connection current measurement N L1 L2 L3 PE I1 I2 I3 IN I5 I6 Voltage measurement terms connections Rogowski coil RCM...

Страница 8: ...rect setting of the current transformer check the performance values of the individual phases these must not possess any negative signs when drawing energy Also check the power factor for plausibility...

Страница 9: ...Power supply 230 V 50 Hz 50 VA Interface Ethernet Measurement inputs See UMG 96RM E operating instructions Measurements See UMG 96RM Eoperating instructions Dimensions L x W x H 350 mm x 295 mm x 150...

Отзывы: