Go-X Series Rolling Shutter GigE Vision I/F (Preliminary)
64
AutoLevelControl
Configure Auto Level Control settings.
Item
Setting range
Default value
Description
ALCControlReference
0: Peak Channel,
1: Selected Channel
1: Selected Channel
ALCControlChannel
0:Red,
1:Green,
2:Blue
1: Green
ALCReference
30
~
95
50
Set the target level for ALC. (unit: %)
ALCAreaSelector
―
Low Right
Select the area for which to configure [ALCAreaEnable].
[Setting range]
0:Low Right,
1:Low Mid-Right,
2:Low Mid-Left,
3:Low Left
4:Mid-Low Right, 5:Mid-Low Mid-Right, 6:Mid-Low Mid-Left, 7:Mid-Low Left
8:Mid-High Right, 9:Mid-High Mid-Right, 10:Mid-High Mid-Left, 11:Mid-High Left
12:High Right,
13:High Mid-Right,
14:High Mid-Left,
15:High Left
ALCAreaEnable
True, False
True
Enable/disable the photometry area selected in [ALCAreaSelector].
ALCAreaEnableAll
True, False
True
True: Operate ALC with all areas designated as photometry areas, regardless of the
individual enabled/disabled photometry area states configured in [ALCAreaSelector].
False: Operate ALC according to the individual enabled/disabled photometry area states
configured in [ALCAreaSelector].
ALCControlRatio
1
~
100
90
Set the response speed. (100 is the fastest.)
AutoShutterControlExposureMin
100
~
100
Set the minimum value for the ExposureAuto(ASC) control range.
Setting Range
Min: 100
Max: AutoShutterControlExposureMax -1
AutoShutterControlExposureMax
―
GOX-6409MC-PGE:
55328
GOX-12409MC-PGE:
105782
GOX-20409MC-PGE:
173308
Set the maximum value for the ExposureAuto(ASC) control range.
Setting Range
Min:
AutoShutterControlExpos1
Max
: ExposureTimeMax(04 – 29) Integer Type Value
AutoGainControlGainRawMin
100
~
100
Set the minimum value for the GainAuto(ASC) control range.
AutoGainControlGainRawMax
Max:
GOX-6409MC-PGE: 12600
GOX-12409MC-PGE: 1600
GOX-20409MC-PGE: 1600
1600
Set the maximum value for the GainAuto(ASC) control range.
ALCStatus
0: Off, 2: ASC, 3: AGC
0:Off
Allows confirmation of the current operation area during ALC operation.
AutoControlStatus
―
Idle
Allows confirmation of the AGC, ASC, and AWB convergence status.
[Status]
1:ExecutingASC, 2:ExecutingAGC, 3:ExecutingASCandAGC, 4:ExecutingAWB
5:ExecutingASCandAWB, 6:ExecutingAGCandAWB, 7:ExecutingASCandAGCandAWB,
8:Convergent, 9:ConditionError, 255:Idle
Содержание GOX-12409C-PGE
Страница 72: ...Go X Series Rolling Shutter GigE Vision I F Preliminary 72 Spectral Response GOX 6409MC PGE ...
Страница 73: ...Go X Series Rolling Shutter GigE Vision I F Preliminary 73 Spectral Response GOX 12409MC PGE ...
Страница 74: ...Go X Series Rolling Shutter GigE Vision I F Preliminary 74 Spectral Response GOX 20409MC PGE ...
Страница 75: ...Go X Series Rolling Shutter GigE Vision I F Preliminary 75 Dimensions ...