Chapter 5
– Setup Menu
56
If measurements are stored in RAM or on an SD card, they can be accessed and recalled.
In the preceding example, measurement number 2 of 3 is displayed.
Y
X
,
Y
(
OR X
,
Y
)
D
ATA
E
NTRY
P
ROCEDURE
Since the PR-7XX measures the spectral content of the sample, it is possible for the instrument to calibrate the
display white point based on user entered values for luminance and CIE chromaticity . The user can select to
enter
Y x,y
(luminance and CIE 1931 x, y values) or
x,y.
If
Y x,y
is selected, after calibration of the display, the
luminance is properly adjusted to the entered value for
Y
. If
x,y
is chosen, the absolute luminance value is
ignored and the display is calibrated based on chromaticity values only.
FIGURE 49 - Y XY DATA ENTRY SCREEN
Y
X
,
Y
P
ROCEDURE
1.
Touch the ▲▼icons adjacent to
Mode
to select
Match Data.