Chapter 4
– Basic Setup Procedures
42
R
ESET TO
D
EFAULT
Resets all
Preferences
to their factory default settings. Factory values are:
Parameter
Setting
Power Saving
Enabled
Brightness and
Contrast
Brightness: 50%
Contrast: 50%
Brightness during measure: 25%
Sounds and Alerts
All Enabled
Connectivity
High Power
Date and Time
US Pac. Coast Date and Time
Units
Measurement Units: Metric
Lum. Intensity: Candelas
RS-232 Connectivity
Baud Rate: 115200
Setting the Default Parameters:
1.
Touch
Reset to Default
from the
Preferences
menu. The following screen appears.
2.
Touch
OK
to reset all Preferences or
Cancel
to exit without making changes.
H
ARD
R
ESET
If the instrument ever becomes unresponsive, perform a hard reset. Press and hold the Power Off (O) and
Measure button simultaneously until the instrument shuts itself off. Upon rebooting, the instrument should
function normally. If not, contact Photo Research for assistance.
Содержание Photo Research SpectraScan PR-7 Series
Страница 2: ......
Страница 10: ...PR 7XX User s Manual vi...
Страница 19: ...Chapter 2 System Overview 15 To traverse back to the System Menu simply touch the Back button twice...
Страница 32: ......
Страница 84: ......
Страница 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Страница 92: ......
Страница 102: ......
Страница 108: ......
Страница 117: ...Chapter 12 Connectivity 113...
Страница 118: ......