UMMARY
The following table summarizes Data Codes and their meanings. For full details, please see the Expanded Code
table following.
Data Code
Description
0
Repeat last response code
1
status, units, Photometric brightness, CIE 1931 x,y
2
status, units, CIE 1931 Tristimulus Values (X, Y, Z)
3
status, units, Photometric brightness, CIE 1976 u’, v’
4
status, units, Photometric brightness, Correlated Color Temperature, Deviation from
Plancks Locus in 1960 u,v units
5
status, units, Peak Wavelength, Integrated Power, Integrated Photon, WL, Spectral
Data at each WL
6
status, units, Photometric brightness, CIE 1931 x, y, CIE 1976 u’, v’
7
status, units, Photometric brightness, CIE 1960 u,v
8
status, Raw (uncorrected) light per pixel
9
status, Raw (uncorrected) Dark Current per pixel
10
status, Raw Light minus Dark Current per pixel
11
status, units, Scotopic Brightness
12
status, units, Photometric brightness, CIE 1931 x, y, CIE 1960 u, v
13
status, Gain description, exposure time in milliseconds
14
status, Sync mode description, sync period in milliseconds
15
Status, bandwidth used for last measurement
110
status, Instrument Serial Number
111
status, Instrument Name
112
status, Number of Accessories, Number of Apertures
114
status, Software Version
115
status, Battery status
116
status, Accessory List
117
status, Aperture List
118
status, Bandwidth list
120
status, Hardware configuration
200
status, Last Meas. Max Raw Light Value, Last Meas. Min Raw Light Value, Last Meas.
Avg Raw Light Value.
201
status, Same as report 200 for Raw Dark
400
status, Contents of Last Measurement buffer.
401
status, Number of measurements stored in RAM
402
status, Directory of measurements stored in RAM. An error message is generated if
there are no stored measurements.
411
status, List of files in SD card and number of stored measurements per file.
412, filename
Status, Directory of stored measurements in the file “filename” in SD card.
601
status, Current Setup Report
– comma delimited
602
status, Current Setup Report, with labels.
TABLE 8 - DATA CODE SUMMARY
Содержание Photo Research SpectraScan PR-7 Series
Страница 2: ......
Страница 10: ...PR 7XX User s Manual vi...
Страница 19: ...Chapter 2 System Overview 15 To traverse back to the System Menu simply touch the Back button twice...
Страница 32: ......
Страница 84: ......
Страница 91: ...Chapter 7 File Settings 87 FIGURE 73 ILLUMINANCE LUMINANCE CONCEPT...
Страница 92: ......
Страница 102: ......
Страница 108: ......
Страница 117: ...Chapter 12 Connectivity 113...
Страница 118: ......