PhaseStation 53100A User’s Manual V1.04 25-Apr-2022
46
inconsistent drift characteristics, its short-term ADEV will eventually resemble the HDEV
performance that was observed only a few minutes after power-up.
Time Deviation (
t
)
TDEV, or
σ
x
(τ
), expresses the time stability of phase at the specified tau, in units of
seconds. Numerically, TDEV is equal to (MDEV *
τ
) / sqrt(3). It’s similar to the TIE (Time
Interval Error) statistic used by the telecommunications sector.
Maximum Time Interval Error (
M
)
Simply put, MTIE represents the worst-case timing performance of the clock under test over the
tau intervals covered by the measurement, as observed by sliding windows of the corresponding
tau durations through the phase record. Like TDEV, MTIE is frequently encountered in
telecommunications standards specifications.
Along with the other stability metrics supported by TimeLab, the MTIE plot is computed on an
incremental basis as successive data points arrive from the counter or phase-difference analyzer
and updated in real time. Because MTIE calculations are more time-consuming than the other
stability metrics supported by TimeLab, they are performed on only the first 10,000 points of the
phase record by default. This limitation can be removed by deselecting the
File→Limit MTIE
processing to 10K points
option.
Содержание PHASE STATION 53100A
Страница 1: ...TM 53100A Phase Noise Test Set User s Manual Revision 1 04 April 25 2022...
Страница 16: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 16...
Страница 18: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 18...
Страница 20: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 20...
Страница 30: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 30...
Страница 36: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 36...
Страница 42: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 42...
Страница 64: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 64...
Страница 74: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 74...
Страница 104: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 104...
Страница 116: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 116...
Страница 122: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 122...
Страница 130: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 130...
Страница 140: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 140...
Страница 143: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 143 Figure 66 5 MHz residual phase noise and AM noise...
Страница 214: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 214...
Страница 215: ...PhaseStation 53100A User s Manual V1 04 25 Apr 2022 215 Appendix Legal and regulatory notices...