M-2000® Hardware Manual
8.2
Field of View Calibration
Note:
This is only necessary once unless the configuration files have been
lost (done at the factory setup). So, the “point and click” function
can simply be verified before performing the calibration.
1.
The previous section (Marking the Measurement Beam Position) should
be completed before proceeding.
2.
Select
Hardware
>
Controls
>
Misc
>‘Additional Camera Functions’ to open
the Camera Functions Window. Select ‘Measure Camera Field of View’.
Figure 8-18. Additional Camera Functions button – Measure Camera Field Of View.
3.
Adjust the illumination, if applicable, so the sample surface is visible.