Wobble Correction
In order to measure without aligning at each rotator (theta) angle, the wobble
must be removed from the sample stage.
1.
Mount and align the sample (refer to Section
Error! Reference source not
found.
Error! Reference source not found.
) with the theta position at 0°.
2.
Select
Hardware
>
Controls
>
System
>‘Translator’ (or ‘Rotator’) and move
the theta position to 180°.
Figure 7-19. Translator Position dialog.
3.
Adjust the Sample Tilt to move the Lookdown cross hair position half of
the distance it has changed.
Figure 7-20. Sample Tilt adjustment.
4.
For example, if X=8, Y=-6.5 : adjust so X=4, Y=-3.25.