High Voltage Probe HV-P60A
3. Adjustment of Probe
A waveform adjustment is necessary for this probe. A waveform
adjustment of this probe adjusts the capacity of the variable capacitor
that exists in this probe as the gain steadies for the frequency when
the oscilloscope is combined with this probe. If this adjustment is not
appropriately done, a correct measurement cannot be done.
Adjust the waveform of this probe when this probe is connected and
used for the oscilloscope.
The waveform of this probe is adjusted with the following
preparations according to the procedure.
3.1 Preparations
Prepare the signal generator that can generate the following signal
specifications and the measuring instrument (oscilloscope) used for
the measurement thereafter.
As for the example of the used equipment, it is not the following.
< Signal generator >
・
Waveform
: Square wave
・
Frequency
: 1 kHz or less
・
Output voltage : 50 V to 100 V
・
Rise time
: 1
μ
s or less
・
Example of the used equipment
Example: KHT-1000 (PMK)
< Oscilloscope >
・
Example of the used equipment
Example 1: DS-5400A series
(IWATSU TEST INSTRUMENTS CORPORATION)
Example 2: DS-5100B series
(IWATSU TEST INSTRUMENTS CORPORATION)