
ManagerIQEasy_UM_9752103002_GB_V3_1
118
6.30 Calibrating a bar (Expert user)
If it is necessary to calibrate a bar, follow the following procedure. To be able to calibrate a bar,
the user must be logged in as an Expert (see Section 6.8).
If necessary, go to the main screen
(Home).
If necessary, log in as
(User level, Expert), see Section 6.9.
Press
or
(Device).
If necessary, press
(Maintenance) to go to the Maintenance screen.
Browse using the
and
buttons to the page with the Action “
Calibrate bar
”
(You can also use the large non-visible areas to browse).
Press
(Action) after Action “
Calibrate bar
”.
Within a few seconds, after the Action button the message “
Calibration started...
” will be displayed,
and shortly thereafter this will be followed by the message “
Calibration in progress (10%)...
”.
The counter will slowly increase until the bar calibration is complete. The text will disappear
when the calibration is complete. In the “
Last calibration date
” parameter the current date and
time will be saved.
Note:
- Sometimes the counter stays at 70%. In this case, disconnect the Device, wait about 10 seconds,
and reconnect the Device. Check if necessary whether the current date and time have been
saved in the parameter “
Last calibration date
”. In that case, the bar is correctly calibrated.
Press
(Home) to return to the main screen and log in again as a Basic user (see Section 6.9).