Harmonic Measurement Commands
Copyright © Itech Electronic Co., Ltd.
35
Returned value
<NR2>
MEASure[:SCALar]:HARMonics:CURRent:FUNDament
al?
This command will trigger the measurement operation to get a real-time current
fundamental harmonic.
Command Syntax
MEASure[:SCALar]:HARMonics:CURRent:FUNDamental?
Returned value
<NR2>
FETCh[:SCALar]:HARMonics:CURRent:THARmonic?
This command is used to read the latest pre-processing total harmonic
current .It will not affect the device setting values and will not trigger a new
measurement operation.It only requires to get the latest available value that has
been measured.
Command Syntax
FETCh[:SCALar]:HARMonics:CURRent:THARmonic?
Returned value
<NR2>
MEASure[:SCALar]:HARMonics:CURRent:THARmonic
?
This command will trigger the measurement operation to get a real-time total
harmonic current.
Command Syntax
MEASure[:SCALar]:HARMonics:CURRent:THARmonic?
Returned value
<NR2>
FETCh[:SCALar]:HARMonics:CURRent:THDistort?
This command is used to read the latest pre-processing total harmonic
distortion factor of current.It will not affect the device setting values and will not
trigger a new measurement operation.It only requires to get the latest available
value that has been measured.
Command Syntax
FETCh[:SCALar]:HARMonics:CURRent:THDistort?