Current Measurement Commands
Copyright © Itech Electronic Co., Ltd.
14
Chapter3 Current Measurement Commands
This chapter describes the current relevant commands.
FETCh?
This command is used to read all relevant parameters of current, voltage and
power measured by the power meter. The order in which the parameters are
read is as follow.
VOLT_RMS, VOLT_MN, VOLT_RMN, VOLT_DC, VOLT_AC, VOLT_MAXPk,
VOLT_MINPk, VOLT_PPEak, VOLT_CF, FREQ_VOLT,
CURR_RMS,
CURR_MN,
CURR_RMN,
CURR_DC,
CURR_AC,
CURR_MAXPk, CURR_MINPk, CURR_PPEak, CURR_CF, FREQ_CURR,
CURR_INR,
POWER_ACTive, POWER_REACtive, POWER_APParent, POWER_PF,
POWER_Phase, FREQ_SSOurce
Please refer to IT9120 User Manual for the detailed instructions of the
parameters.
Command syntax:
FETCh?
Returned value
<NR2>, <NR2>, ......, <NR2>
FETCh[:SCALar]:CURRent:AC?
This command is used to read the latest pre-processing current.It will not affect
the device setting values and will not trigger a new measurement operation.It
only requires to get the latest available value that has been measured.
Command syntax:
FETCh[:SCALar]:CURRent:AC?
Returned value
<NR2>
MEASure[:SCALar]:CURRent:AC?
This command will trigger the measurement operation to get a real-time current.
Command syntax:
MEASure[:SCALar]:CURRent:AC?
Returned value:
<NR2>
FETCh[:SCALar]:CURRent:CFACtor?
This command is used to read the latest pre-processing current crest factor.It