List parameters description
:
Parameter
Description
Voltage AC
Voltage value and slope.
Start Phase
Start phase setting
By default, the program automatically calculates
this value (which is guaranteed to be continuous
with the previous step). If the user modifies this
parameter, the user set value is used without
guarantee of continuity.
Phase Difference
Phase difference between ABC, only displays in 3-
phase mode.
Frequency
Frequency and slope.
Waveform
Waveform type, every basic waveform can be
selected.
Step jump
Method of step jumps to next step.
Time: when the time is out, jumps to next step
Trig: reveive a trigger signal, jumps to next step.
Phase: jumps to next step at this phase.
Trig out
Whether outputs a signal when this step is ending.
4.
Press
[Esc]
to return Edit interface, Press
[Save].
At list Edit interface, click the step number, the
[Insert]/[Paste]/[Cut]/[Copy]
/[Delete]
will display, click the key to edit.
5.
Press [
Esc
] to return.
7.1.2 Select/Run List File
If several List files are edited, press Recall to recall the List file to be tested.
Detailed operation steps are as below:
1.
Press
[Shift]+[V-set](list
) on the front panel to enter the List function
configuration interface.
2.
Press
[Open]
, select the saved List01.csv file, and press
[Enter]