8.0   COMPONENT REPLACEMENT

The procedures for checking and replacing components of the HEMOCHRON® Jr. Signature
microcoagulation instrument are described in the following sections. 

8.1 Examine the Fuses and Battery Pack Connections 

Procedure:

8.1.1 Position the circuit board as shown in 

Figure 3

.

8.1.2 Measure the resistance across fuse F1 and fuse F2 to verify that neither is

blown. If the resistance across a fuse is greater than 100 ohms, arrange to
have the instrument returned and have ITC service personnel replace the
circuit board and to have the time and date reset. 

8.2 Replace the Battery Pack (If Necessary)

Procedure:

8.2.1 Follow

7.0 DISASSEMBLY PROCEDURES.

8.2.2 Position the HEMOCHRON

®

Jr. Signature  microcoagulation instrument base

so the battery is facing you as shown in 

Figure 5

.

8.2.3 Using a Phillips screwdriver, remove the two battery mounting screws.

Remove the battery.

8.2.4 Replace the battery with a new battery.

Note: Dispose of the Nickel-Cadnium (NiCad) battery in accordance with local

regulations. 

8.2.5 Reassemble instrument per section 9.

8.2.6 Charge the battery 12 to 18 hours before use.

8.2.7 Reset the time and date for all models except JrII. 

8.3 Inspect the Pump Assemblies (If Necessary)

If the sample and waste channels of the test cuvette were not properly filled (see
the Diagnostic Procedures section of this manual) but the motors were found to
work, the pump assemblies should be inspected.

Procedure:

8.3.1 Position the HEMOCHRON

®

Jr. Signature  microcoagulation instrument base

so the battery is facing you as shown in 

Figure 5

.

8.3.2 Using a Phillips screwdriver, remove the screws that secure the two mounting

brackets. Remove the mounting brackets but do not remove the pump
assemblies.

8.3.3 Carefully visually  inspect each pump assembly, seals, and tubes for leaks. 

8.4 Replace the Pump Assembly (If Necessary)

Note that the replacement pump assembly is pre-tested and pre-torqued and 
therefore no adjustment is required.

HEMOCHRON

®

Jr. Signature Service Manual

10

Содержание HEMOCHRON Jr. Signature

Страница 1: ...SERVICE MANUAL...

Страница 2: ...Replacement Parts List 3 0 Specialized Test Equipment 4 0 Troubleshooting 5 0 Diagnostic Procedures 6 0 Disassembly Procedures 7 0 Component Replacement 8 0 Reassembly 9 0 System Verification 10 0 Fi...

Страница 3: ...the membrane switch panel For a complete listing of hazard and fault messages reference the Operator s Manual This manual does not cover the diagnostics or replacement of the printed circuit board as...

Страница 4: ...the instrument it should be cleaned and disinfected using the procedures approved by your laboratory to avoid spreading contamination to laboratory personnel or equipment A 10 bleach water solution i...

Страница 5: ...rt Number Version HJ6015X Hemochron Jr Signature HJ6015X Hemochron Jr Signature International FA5509X Hemochron Jr II IL6015X GEM PCL HV6015X SCA 2000 NOTE The versions listed are covered by this manu...

Страница 6: ...aspirated by may need to be replaced SMALL STS the instrument PREMATURE Sample was detected at front Repeat test with new cuvette 6 2 7 0 8 3 8 4 SAMPLE detector before specified If message persists a...

Страница 7: ...es Add Sample and Press Start are displayed the batteries and fuses are OK If the battery fuse problem persists check the fuse and or remove and replace the battery per sections 7 0 8 1 8 2 9 0 and 10...

Страница 8: ...ll switches and monitoring for the correct display For this test you will need an unused test cuvette See Figure 1 This procedure is not applicable to Jr II Procedure 6 3 1 Insert the test cuvette wit...

Страница 9: ...board Release the tube fitting by sliding it to the left as shown in Figure 3 7 2 3 Release the second tube fitting in the same manner The front of the circuit board can now be lifted further Importan...

Страница 10: ...place the battery with a new battery Note Dispose of the Nickel Cadnium NiCad battery in accordance with local regulations 8 2 5 Reassemble instrument per section 9 8 2 6 Charge the battery 12 to 18 h...

Страница 11: ...ecially careful that the waste pump assembly does not interfere with the battery pack 8 4 9 Connect the new pump tube to the fitting 8 4 10 Connect the motor connector to the circuit board as describe...

Страница 12: ...fitting into the other circuit board slot Make sure that the tubes are not kinked 9 2 4 Check that all of the leads are connected to the circuit board and reconnect the Keypad Overlay Top Cover Assemb...

Страница 13: ...es and a HEMOCHRON Jr Signature Microcoagulation Whole Blood Quality Control kit The procedures for performing the verification are described below 10 1 Run an Electronic System Verification Test Proc...

Страница 14: ...more times using a new test cuvette for each replicate 10 2 5 Examine the results that were obtained If no fault messages were displayed and if the results on the replicates are within the Acceptable...

Страница 15: ...11 0 FIGURES HEMOCHRON Jr Signature Service Manual 15...

Страница 16: ...HEMOCHRON Jr Signature Service Manual 16 FIGURE 1...

Страница 17: ...RETENTION SCREWS HEMOCHRON Jr Signature Service Manual 17 FIGURE 2 CUVETTE HOUSING MOUNTING SCREWS...

Страница 18: ...FUSES F1 F2 BATTERY LEAD CONNECTOR TUBE FITTINGS HEMOCHRON Jr Signature Service Manual 18 FIGURE 3 MOTOR CONNECTOR SAMPLE J5 MOTOR CONNECTOR WASTE J4 LED HARNESS ASSY CONNECTOR J1...

Страница 19: ...HEMOCHRON Jr Signature Service Manual 19 FIGURE 4 CONNECTOR J6...

Страница 20: ...HEMOCHRON Jr Signature Service Manual 20 FIGURE 5 BATTERY MOUNTING SCREWS WASTE PUMP ASSEMBLIES SAMPLE MOUNTING BRACKET SCREWS...

Страница 21: ...HEMOCHRON Jr Signature Service Manual 21 FIGURE 6 LED HARNESS PCB MOUNTING SCREWS...

Страница 22: ...ARD ASSEMBLY J6 KEYPAD UPPER SHELL ASSY SEE SECTION 3 J7 BATTERY PACK ASSY HJ6011X J3 CUVETTE SWITCH J2 HEATER J1 IR LED DETECTOR FA5059X J5 SAMPLE MOTOR PUMP ASSY HJ6003X J4 WASTE MOTOR PUMP ASSY HJ6...

Страница 23: ...HEMOCHRON Jr Signature Service Manual 23...

Страница 24: ...HJ7430 1 03 HEMOCHRON is a registered trademark of ITC...

Отзывы: