Page 18 of 29
Fast Cal Series Iss.07 – 08/13
The upper display will show the current value that can be modified with the UP and DOWN keys.
MONITORING THE INDICATOR STATUS
For the indicator the REM beacon is lit continuously when the user calibration is active, the REM beacon flashes when
the PC communications port is active.
Units
Momentary pressing of the Scroll key will show the controller units °C or °F.
ADVANCED INDICATOR OPERATION
The indicator can be configured with up to five custom calibration points; the points contain "data pairs". First the
temperature (point) and secondly the Error (offset) at this temperature point. Isotech Dry Block calibration certificates will
show the values to suit a particular sensor.
These values can be inspected, and modified with the following procedure,
Hold the PAGE key down until the display indicates,
CONF
GOTO
then
0
CODE
Set the Code to 2 with the UP key
The display reads,
CONF
Press the Page Key until the controller shows
CAL
Now use the Scroll key to examine the data pairs. The values Pnt 1- Pnt 5 and Ofs 1 to Ofs 5 can be modified with the
UP and DOWN keys.
To exit this mode hold the Page key until the top display shows,
CONF
GOTO
And then set the upper display to Lev 1. While in this mode take care not to modify other parameters.