Heatpipe Furnace Models 17702W, 17702P, 17702S (13
–
07/20)
Below is the procedure to install a program for Channel 1 (Heatpipe furnace)
To install for The Annealing furnace, ensure Channel 2 is enabled in the settings:
1.
Log in as the Engineer
2.
Scroll to PROGRAMMER and enter
3.
Scroll to SET UP and enter
4.
Scroll to CHANNELS and enable channel 2
To create, install and run a PROGRAM via the controller interface please follow the
instructions below:
See below for a more detailed description of each of the PROGRAM functions.
PROGRAM Details:
Operation
This allows the user to select one of the following:
Load. Opens the program store and allows the user to select a program to be loaded.
Store. Allows the current program to be saved to the internal program drive. This is useful if you wish to
snapshot the current program and store this under a different program name.
Delete. Allows the selected program to be deleted.
Delete All. Deletes all programs.
Copy. Copies the selec
ted program for ‘pasting’ either from the internal drive to the USB device, or vice-
versa. This is useful if you wish to transfer a program to other instruments.
Copy All. As above, for ‘Copy’, but copies all the programs in the selected directory.
Note:
If a ‘Store’, ‘Copy’ or ‘Copy All’ operation would result in there being a total of more than 100
program files in the internal drive, the operation fails and an error message is displayed.
Status
Success. Previous operation was successful.
Failed. Previous operation failed.
Loading. The program is loading.
Copying. The program copy process is underway.
Deleting. The relevant program is being deleted.
Program
The name of the program currently loaded.
Holdback Style
Appears only if ‘Holdback’ is enabled. See ‘Holdback’, below.
Program
Holdback applies to all appropriate segments.
Per Segment: Holdback enabled on a segment by segment basis as described in ‘Segment configuration
below.
Ch1 Holdback
A
ppears only if ‘Holdback Style’ (above) is set to ‘Program’.
Off: Holdback is disabled
Low: Holdback is entered when PV < (PSP - Holdback Value)
High: Holdback is entered when PV > (PSP + Holdback Value)
Band: Holdback is entered when PV < (PSP - Holdback Value) or PV > (PSP + Holdback Value)