 Warning

 – For non-contact voltage sensing, do not 

touch Hazardous Live conductors with the red or black 

test-probe.

•  For non-contact voltage sense, place the top of  

the instrument body or the red test-probe near  

the conductor.

•  The buzzer will sound and the number of dashes 

shown on the LCD will indicate the relative electric field 

strength. If there is no indication, a voltage may still be 

present.

For calibration and repair contact RS Components - the 

address is given at the end of these instructions.

Cleaning

Periodically wipe the case with a dry cloth and detergent.

Do not use abrasives or solvents.

Battery Replacement

 WARNING

Disconnect the test leads from the circuit before 

removing the battery cover.

Refer to the following figure to replace the batteries: 

Specifications

1-1 General Specifications

LCD display digits:

 

4000-count large-scale Liquid Crystal Display.

Measuring rate:

 3 times / sec.

Polarity Indication:

 Automatic, positive implied,

Overrange display:

 “OL” or “-OL” 

Unit symbol indication.

Automatic power off time:

 

Approximately 20 minutes after power on. 

Low battery indicator:

    is displayed. 

Power requirement:

 2 x 1.5V LR44 type batteries.  

Battery life:

 50 hours (GPA76P)

1-2 Environmental Conditions

Indoor Use Only.

Calibration:

 One year calibration cycle.

Operating temperature: 

0°C to 30°C ( 80% RH) 

30°C to 40°C ( 75% RH) 

40°C to 50°C ( 45%RH)

Operating altitude:

 2000m (6562 ft)

Pollution degree:

 2 

EMC:

 EN 61326-1   

Dimensions (W x H x D):

 56 x 12 x 112mm

Weight:

 115g

Accessories:

 Batteries (installed), carry case and 

Instruction manual.

1-3 Electrical Specifications

Accuracy is ±(% r number of digits) at 

23°C  ± 5°C < 80%RH.

Storage temperature: 

-20 to +60°C, 0 to 80% RH (batteries not fitted).

Temperature coefficient:

 Add 0.2 x (Specified 

accuracy) / °C, < 18°C, > 28°C.

Shock and vibration:

 Sinusoidal vibration per 

MIL-T-28800E (5 to 55 Hz, 3g maximum).

Drop Protection:

 4 feet drop onto hardwood or 

concrete floor.

Overvoltage (Measurement) category: 

IEC 61010-1 300V CAT 

, 600V CAT 

.

Measurement Category l

 is for measurements 

performed on circuits not directly connected to mains. 

Examples include: Measurements on battery powered 

equipment and specially protected (internal) mains-

derived circuits.

Measurement Category ll

 is for measurements on 

circuits directly connected to the low voltage installation. 

Examples include: Household appliances, portable tools 

and similar equipment.

Measurement Category lll

 is for measurements 

performed in the building installation. Examples  

include measurements on distribution boards, junction 

boxes, socket-outlets and wiring and cables in the  

fixed installation.

Maintenance

Do not attempt to repair this Instrument. It contains no 

user-serviceable parts. Repair or servicing should only be 

performed by qualified personnel. This instrument should 

be calibrated yearly or more frequently if used in harsh 

conditions or if it is suspected of being inaccurate.

Disconnect

I

SELECT

RANGE

HOLD

AC A /DC A 

Volt Sence

Black

Red

A

SENSE

AUTO

DC

A

Red

V

RANGE

HOLD

SELECT

V

RANGE

HOLD

SELECT

Battery Replacement

1

Carry Case Assembling

1

2

2

RANGE

HOLD

SELECT

Battery Replacement

1

Carry Case Assembling

1

2

2

Function Range

Resolution Accuracy

ACV

400.0mV

0.1mV

± (1.5%+ 5 D)

4.000V

0.001V

± (0.9%+ 5 D)

40.00V

0.01V

400.0V

0.1V

600V

1V

Function Range

Resolution Accuracy

Cap

40.00nF

0.01nF

±(5%+ 0.4nF )

400.0nF

0.1nF

±(2.9%+ 5 D)

4.000µF

0.001µF

40.00µF

0.01µF

400.0µF

0.1µF

Function Range

Resolution Accuracy

ACA

400.0µA

0.1µA

± (1.5%+ 5 D) 

4.000mA

0.001mA

Function Range

Resolution Accuracy

DCA

400.0µA

0.1µA

± (0.9%+ 5 D)

4.000mA

0.001mA

Function Range

Resolution Accuracy

%

0.1 - 99.9% 0.1%

± (0.5%+10 D) *

Function Range

Resolution Accuracy

Hz

40.00Hz

0.01Hz

±(0.3%+ 5 D)

400.0Hz

0.1Hz

4.000KHz 0.001KHz
40.00KHz 0.01KH

Function Range

Resolution Accuracy

DCV

400.0mV

0.1mV

± (0.7%+ 5 D)

4.000V

0.001V

± (0.6%+ 2 D)

40.00V

0.01V

400.0V

0.1V

600V

1V

± (0.7%+ 5 D)

Function Range

Resolution Accuracy

Ohm

400.0Ω

0.1Ω

± (0.9%+ 5 D)

4.000kΩ

0.001kΩ

± (0.9%+ 2 D)

40.00kΩ

0.01kΩ

400.0kΩ

0.1kΩ

4.000MΩ 0.001MΩ

± (1.5%+ 5 D)

40.00MΩ* 0.01MΩ

Frequency Response:

 50 to 500Hz 

AC Conversion Type:

 AC Coupled Average sensing, 

RMS Indication.

Input Impedance:

 10MΩ || <100pF.

Overload protection:

 600V rms.

Open Circuit Voltage: 

0.4V

* Rolling of digits of less than ± 2%

Overload protection:

 600V rms.

CONTINUITY CHECK 

Continuity Threshold:

 Approx. <50Ω

Continuity Indicator:

 2.7KHz Tone Buzzer.

Input Protection:

 600V rms.

Measuring Time:

 <30sec. (400.0µF range),

<10sec. (40.0µF range), <3sec. (other range)

Overload protection:

 600V rms.

*30% ≤ Duty ≤ 70%, 

 Square Wave (5 Hz to 1KHz) 

Sensitivity:

 1.5Vp-p

Overload protection:

 600V rms.

Frequency Response:

 50 to 500Hz 

AC Conversion Type: 

AC Coupled Average Sensing, 

RMS Indicating.

Input Impedance:

 Approx. 3KΩ

Overload protection:

 600V rms.

Input Impedance:

 Approx. 3KΩ

Overload protection:

 600V rms.

 

DIODE TEST

Test Current:

 Typical. 350µA

Open Circuit Voltage:

 Max. 1.8V DC

Input Protection:

 600V rms.

Sensitivity:

 > 10Vp-p (40KHz range) > 1.5Vp-p  

(other ranges) 

The signal must have positive and negative portions  

of a cycle.

Max. Sensitivity:

 600V rms.

Overload protection:

 600V rms.

Input Impedance:

 10MΩ || <100pF.

Overload protection:

 600V rms.