Section 2 Programming
2-5
• Number of Samples - Set the number of
samples to collect or place the GLS in the
Continuous Sampling mode.
• Sample Volume - Enter the desired
volume to collect at each sample event.
• Program Start Time- Without a delay the
GLS will always take the first sample
when you press the Go button. This
programming step allows you to delay
the first sample.
• Suction Line - Enter the type of suction
line in use. This step also includes the
Sample Volume Calibration procedure.
2.4.1 Sample Pacing
The GLS displays two pacing options - Time
Paced and Flow Paced. Select Time to collect
samples at uniform time intervals. Select Flow
to collect samples based on flow volumes. An
input signal from an external instrument is
required when you select Flow Paced.
The current selection will be blinking. To
program the pacing method:
1. Press the Arrow button to select an option.
2. Press the Enter button to accept the
blinking option. The GLS loads the option
into the current program settings and
advances to the next step.
TIME PACED
FLOW PACED