Operation
82
Modline 5
5
Operation
5.1
Sensor Rear Panel Setup and Operation
5.1.1
Introduction
Become familiar with the displays, pushbutton controls, menus and the functions required for Sensor
setup and operation as described below. Modline
®
5 instruments are thoroughly tested and calibrated
to factory specifications before shipment and should require no more than brief function selections and
adjustments to meet the requirements for measurement of process temperatures.
5.2
Rear Panel Keyboard, Displays and Sight and Laser Aiming
5.2.1
Description
Located on the rear of the Modline 5 Sensor are: a keyboard panel with three setup pushbuttons, three
red word indicators for
Setup
,
Alarm
and
Window
(for the Dirty Window Detector optional feature),
and a four character green alphanumeric Light Emitting Diode dot matrix display with adjacent green
°F
and
°C
indicators.
The display is identified as the Temperature (
TEMP)
display in this manual. The TEMP display and
°F
and
°C
indicators are used for temperature indication in Fahrenheit or Celsius.
LOW
or
HIGH
is
displayed when viewing target temperatures below or above the Sensor temperature range.
INV
is
presented on the display for Invalid ratio temperature measurement conditions. Setup parameters will
be shown on the TEMP display during setup and error codes displayed in response to an Alarm inquiry.
Two of the pushbuttons are Down / Up Arrows labeled
▼
and
▲
. The third is an Enter pushbutton
labeled
ENT with the keyboard enter symbol
. These buttons are identified as
▼▲
and
in this
manual.
A sight for visually aiming the Sensor is located in the center of the keyboard. The process target and
Sensor circular reticle can be seen when viewing through the sight. Sensors with the Laser aiming option
have a pushbutton to energize a laser in place of the sight.
Access to the keypad requires removal of a screw on protective window with O-ring gasket. The
protective window must be in place during temperature measurement to maintain the NEMA 4 (IP65)
Sensor housing rating.
Figure 74: Sensor with Visual Through the Lens Sight
Temperature in °C or °F
LOW
=
Below zero scale temperature
HIGH
= Above full scale temperature
INV
= Invalid Condition for Ratio
Temperature Measurement
Содержание Modline 5
Страница 2: ......
Страница 4: ......
Страница 156: ...Notices 154 Rev L6 Feb 2017 Modline 5 9 Notices ...