Wireless Digital Wireless Digital Flat Panel Detector Mars1417X
5.3.3.
IDLE
XWIN1
(fixed)
X-Ray
imaging
XWIN1
(fixed)
Offset
imaging
X-Ray
Info
RAW Preview
RAW X-Ray image
(
If HW-offset is disabled
)
RAW Offset(Or corrected)
image
1.
For Freesync mode, there is no low power state.
2.
When the detector is Idle, user can start the exposure flow any time.
3.
When the X-Ray starts, the detector will sense the X-Ray automaticlly, the XWIN is configured
by parameter “Set Delay Time” on “Detector” page “Parameter” tab, the unit is “ms”, user
needs to make sure that the XWIN is larger than the X-Ray time.
4.
After the XWIN is end, then the detector will start the acquisition flow.
5.
The preview image will be always sent, which is 4x4 averaging, the raw X-Ray image will be
sent if the HW correction is disabled with the raw offset image follows, otherwise, the X-Ray
image will not be sent and only the corrected image will be transferred.
5.4.
After use
1. Disconnect the software
2. Power off
3. Keep it clean
4. Store under specified conditions
5.5.
Firmware Upgrade
On “Detector” Page
,“
Parameter” Tab
,
user can upgrade firmware by entrance button “Upgrade
Firmware”.
58
User Manual of Mars1417X
Содержание Mars1417X
Страница 2: ......
Страница 6: ...Wireless Digital Flat Panel Detector Mars1417X Symbols and Conventions 4 User Manual of Mars1417X ...
Страница 32: ...Wireless Digital Flat Panel Detector Mars1417X Gigabit Ethernet Cable 1pcs 2 7 1 Detector iRay Technology Co ...
Страница 69: ...5 Operation Instructions for Image Acquisition User Manual of Mars1417X 67 ...
Страница 78: ...Flat Panel Detector Mars1717X 6 6 Battery Charger Label 62 User Manual of Mars1417X ...
Страница 79: ...6 Regulatory Information Battery Label User Manual of Mars1417X 77 ...