Installation and Checkout
This chapter explains how to install and check out the monitor.
2.1
Preinstallation
Refer to Chapter 13 for a description of facility environmental and power requirements,
dimensions, weight and other specifications for the monitor.
2.2
How to Connect the Escort M8 Vital Signs Monitor to Power
To connect the Escort M8 Vital Signs Monitor to power, perform the following procedure:
1.
Plug the power cord into a hospital grade AC outlet.
2.
Plug the other end of the power cord into the power connector on the back of the
Escort M8 Vital Signs Monitor.
3.
Observe that the green Charge LED is now illuminated.
4.
Flip the power switch on the back side of the Escort M8 Vital Signs Monitor to the
ON position.
5.
Observe that the green POWER LED is now illuminated and that the monitor
screen lights up and then goes though the power up routine.
6.
The waveforms and parameter values are not displayed until the Escort M8 Vital
Signs Monitor is connected to a patient.
To power-down the Escort M8 Vital Signs Monitor, perform the following procedure:
1.
Flip the power switch on the back side of the Escort M8 Vital Signs Monitor to the
OFF position.
2.
Observe that the green POWER LED is no longer illuminated and that the monitor
screen goes blank.
WARNING -
For safety reasons it is important that the Escort M8 Vital Signs
Monitor be powered from an outlet where the ground plug receptacle is connected to
earth ground. Do not remove the grounding conductor from the power plug or use
electrical power without proper grounding.
NOTE -
Upon monitor power-on, it may take as long as 30 seconds before the monitor
display is visible.
NOTE -
The monitor should remain connected to the hospital grade AC outlet even when the
Escort M8 Vital Signs Monitor is not in use as this allows the battery to be recharged. The
battery charge indicator will remain illuminated under these conditions.
Содержание Escort M8
Страница 1: ......
Страница 2: ......
Страница 26: ......
Страница 32: ......
Страница 34: ......
Страница 44: ......
Страница 55: ...Repair Revised October 2004 3800 Vital Signs Monitor Service Manual 7 11 Figure 7 10 Input Panel Assembly View...
Страница 56: ...Repair 7 12 3800 Vital Signs Monitor Service Manual Revised October 2004 Figure 7 11 Recorder Subassembly View...
Страница 60: ......
Страница 66: ......
Страница 72: ......
Страница 78: ......
Страница 80: ......