• The device must be connected to a mains socket outlet with protective earthing connections.
• Ventilation requirements: room ventilation.
Installing the instrument
The product may be installed only under the conditions and in the positions specified by Thermo Fisher
Scientific.
Following are the required operating position and conditions:
• Do not place the product in an area where it will be subject to vibration.
• Do not place the product on surfaces, vehicles, cabinets or tables that for reasons of weight or
stability are unsuitable for this purpose.
• Do not place the product on heat-generating surface or near heat emitting devices such equipment
racks or heaters. Verify that there is sufficient clearance between the product and any other system
that may exhaust warm air.
• Do not place the product in an area where its ventilation system is obstructed. Lack of proper
ventilation may result in electric shock, fire and/or serious personal injury or death.
• The product is for indoor use only
• Use only with suitably rated mains supply cord (having 3 conductors min. 16 AWG or 1.5 mm
2
, min.
300V, Harmonized Type for Europe and UL Listed/CSA Certified for North America, with molded
plug rated min. 10A).
• A tolerance of ±10 % shall apply to the nominal input voltage and ±3 Hz to the nominal frequency,
overvoltage category 2.
• Maximum operating altitude 2000 m asl, Maximum transport altitude 4500 m asl.
Explanation of symbols and warnings
Symbol and description
CAUTION!
Risk of danger. Consult the manual for further safety information.
CAUTION!
Risk of electrical shock.
CAUTION!
Do not stare into beam
Use eye protection during servicing
CAUTION!
Potential biohazard.
Appendix H
Safety
Installing the instrument
H
E
‑
Gel
™
Power Snap Plus Electrophoresis System User Guide