General system cleaning
• Rinse the waste tank and funnel tray. The trays are resistant to chemicals typically used in a
western blot, but some of the components of blocking buffers (nonfat dry milk, BSA, etc.) can form
deposits on the surface.
• If needed, use water and a lab tissue to clean the display screen protector, work area window, and
any interior surfaces of the instrument. Using alcohol is not recommended as it may fog plastic
surfaces over time.
Instrument dimensions and specifications
Dimensions
60 cm × 39 cm × 42 cm
Weight
15 kg (32 lbs)
Capacity
4 mini blots, 2 midi blots
Rocks per minute
5-40 rpm
Maximum washes per step
10
Minimum reagent per blot
3.5 mL (mini), 7.5 mL (midi)
Maximum number of protocol steps
9
Maximum number of protocols
7
Buffer dispensing volume
10-150 mL
Incubation step time range
0.1 sec to 96 hrs
Power requirements
12 VDC, 5 A
Relative humidity
5-90%
Operating temperature
4-40°C
Altitude
2,000 m
Storage temperature
-40 to 50°C
Troubleshooting
Observation
Possible cause
Recommended action
Processor won't start.
Power cord plug is inserted in
an inactive wall outlet and/or
not fully inserted in the socket
on the back of the processor.
Ensure the power cord plug is inserted in a live
wall outlet and/or fully inserted in the socket
on the back of the processor.
Chapter 1
Bandmate
™
Automated Western Processor
Troubleshooting
1
12
Bandmate
™
Automated Western Blot Processor User Guide