InMode Without Vasculaze Operator Manual
Section 5: System Operation
36 of 96
Section 5: System Operation
This section of the manual explains how to start the device, operate it, and turn it off.
Prior to using or connecting the Handpiece, inspect the System and
Handpiece for possible mechanical damage.
2.1
Device Start-Up
1
.
Connect the Handpiece to the Handpiece Connector Socket on the System.
2
.
Turn on the Main Power Switch on the Rear Panel.
3
.
Press the On-Off button on the Control Panel to turn the device on. The
System loads the software and enters the Menu Screen.
4
.
Enter a unique password to get access to the device. If the password is correct,
the System enters the Menu Screen.
5
.
The System loads the software and enters a self-test mode. If any problem is
detected during the test the error message will appear (See the
Troubleshooting section in this manual). If the test is passed correctly, then the
System automatically enters the Menu Screen.
6
.
Select Application from the Menu Screen and the System will enter the
Treatment Screen.
1
.
Verify on the screen that the software version is properly displayed, and that
the connected Handpiece type is recognized correctly.
8
.
Select the treatment parameters using Up and Down arrow keys.
1
.
Apply the Handpiece to the treated area, ensuring a full contact.
11
.
Press the Standby icon that will change to Ready.
11
.
To start treatment, press the Footswitch for RF applicators, press the Hand-
Trigger button for IPL applicators, and press the Footswitch followed by the
Hand-Trigger press for laser applicators.
The following information is related to the Diolaze and DiolazeXL lasers only.
12
.
The Cooling Down icon may appear and when the target temperature is
achieved, the icon will change to Standby. If cooling is fast enough, the
Standby icon will appear immediately.
–
If the laser is not activated within 3 minutes, the System will go back to
Standby mode.
–
After pressing the Footswitch, the icon will change to Armed.