InMode Without Vasculaze Operator Manual
Section 2: Safety
6 of 96
Move the System slowly and carefully. The System weighs approximately 32kg
(70lb.) and may cause injury if proper care is not used when moving it.
For RF Handpieces provide as much distance as possible between the system and
other electronic equipment as the activated RF generator may cause interference
between them.
1.2
Fire Hazards
The absorption of diode laser or IPL energy raises the temperature of the
absorbing material. Do not use the System in the presence of explosive or
flammable materials.
Materials conducting RF energy causes temperature rise of the absorbing
material. Do not use the System in the presence of explosive or flammable
materials conductive to RF.
Keep drapes and towels moist to prevent them from igniting and burning. Use
non-flammable prepping solutions.
Do not use flammable substances when preparing the skin for treatment. Be
especially careful with the use of oxygen.
If alcohol is used for cleaning and disinfecting, it must be allowed to dry
thoroughly before the System is used.
1.2
Safety Features of the System
The system incorporates the following safety features. All personnel operating the
system should be familiar with these features.
The System has a unique password to avoid device operation by non-authorized
personnel.
The power electronics cannot be fully activated unless the Applicator and
Footswitch have been connected to the System. With IPL, connection of the
Applicator is sufficient.
An audible tone indicates energy activation.
During activation, the System performs a self-test of the hardware.
The hardware is tested every 10ms to ensure proper operation of electrical
circuit.
The System starts at a low power setting.