DEALER: Keep this manual. The procedures in this manual MUST be 

performed by a qualified technician.

For more information regarding 

Invacare products,  parts, and services, 

please visit www.invacare.com

XPO

2

Portable Oxygen 

Concentrator

Models XPO100 and XPO100B 

Содержание XPO100

Страница 1: ...nual The procedures in this manual MUST be performed by a qualified technician For more information regarding Invacare products parts and services please visit www invacare com XPO2 Portable Oxygen Concentrator Models XPO100 and XPO100B ...

Страница 2: ...ABLE TO UNDERSTAND THE WARNINGS CAUTIONS OR INSTRUCTIONS CONTACT A HEALTHCARE PROFESSIONAL DEALER OR TECHNICAL PERSONNEL BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY OCCUR ACCESSORIES WARNING INVACARE PRODUCTS ARE SPECIFICALLY DESIGNED AND MANUFACTURED FOR USE IN CONJUNCTION WITH INVACARE ACCESSORIES ACCESSORIES DESIGNED BY OTHER MANUFACTURERS HAVE NOT BEEN TESTED BY INVA...

Страница 3: ...To Reduce The Risk Of Burns Electrocution Fire Or Injury To Persons 7 Radio Frequency Interference 8 SECTION 2 COMPONENT REPLACEMENT 9 Replacing the Gross Particle Filter or Filter Cover 9 Accessing XPO2 9 Replacing the Inlet Filter 11 Replacing the Battery 12 Re assembling the XPO2 13 Replacement Parts 13 SECTION 3 CHECKING O2 PURITY 14 SECTION 4 ERROR CODES 15 LIMITED WARRANTY 19 ...

Страница 4: ...that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury or death CAUTION Caution Federal law restricts this device to sale or rental by or on order of a physician or any other practitioner licensed by the law of the State in which he she practices to use ...

Страница 5: ...240 VAC 50 60 Hertz DC Power Supply 11 16 VDC Rated Current Input 1 0 amps at 120 VAC 3 3 amps at 18 VDC Sound Level setting 2 45dBA weighted Altitude Up to 10 000 ft 3046 m above sea level Oxygen Concentration Based on an atmospheric pressure of 14 7 psi 101 kPa at 70 F 21 C 87 to 95 6 after initial warm up period approximately 5 minutes Oxygen Flow Pulse flow delivery Bolus Volume Ranges From 30...

Страница 6: ...mperature Range All power sources Operating temperature 41 F to 95 F 5 C to 35 C Storage temperature 2 F to 140 F 20 C to 60 C Extended Temperature Range Using AC or DC adapters Operating Temperature 95 F to 104 F 35 C to 40 C Continuous use all settings Extended Temperature Range Using Internal Battery Operating Temperature 95 F to 104 F 35 C to 40 C Continuous use settings 1 2 and 3 45 minutes m...

Страница 7: ...cution Fire Or Injury To Persons DO NOT come in contact with the concentrator while wet DO NOT place or store product where it can drop into water or other liquid DO NOT reach for product that has fallen into water Unplug IMMEDIATELY Avoid creation of any spark near medical oxygen equipment This includes sparks from static electricity created by any type of friction NEVER drop or insert any object...

Страница 8: ...erence from even the low levels of electromagnetic emissions permitted by the above standards To determine if the emissions from the XPO2 are causing the interference turn the XPO2 Off If the interference with the other device s stops then the XPO2 is causing the interference In such rare cases interference may be reduced or corrected by one of the following measures Reposition relocate or increas...

Страница 9: ...the filter and inspect for fraying crumbling tears and holes 5 Replace filter cover and or filter 6 Reinstall the filter cover by placing the tabs in the slots and engaging the magnet NOTE Use only Invacare gross particle filter L FIGURE 2 1 Replacing the Gross Particle Filter or Filter Cover Accessing XPO2 WARNING When accessing the XPO2 remember that battery power is still being applied to the i...

Страница 10: ... FIGURE 2 2 Remove Install label 2 Position the XPO2 so that the back panel label is facing up Detail A in FIGURE 2 3 3 Remove the four mounting screws from the back of the XPO2 using a T15 Torx screwdriver Detail A in FIGURE 2 3 FIGURE 2 3 Removing the Front Cover Label XPO2 Mounting Screws Mounting Screws Ribbon Cable Front Cover DETAIL A DETAIL B Back Panel Label Tubing Inlet Filter DETAIL C Fr...

Страница 11: ...etail B and Detail C in FIGURE 2 3 Replacing the Inlet Filter NOTE For this procedure refer to FIGURE 2 4 and FIGURE 2 2 on page 10 NOTE The inlet filter should be replaced as needed 1 Perform the steps in Accessing XPO2 on page 9 2 Remove the inlet filter from the holder Detail A 3 Disconnect the inlet filter with 5 8 inch tubing attached from the 111 2 inch long tubing Detail B FIGURE 2 4 Replac...

Страница 12: ...ing an alligator clip Electrical conduit or a water pipe is normally sufficient when a known good ground in unavailable Care should be taken to ensure that the alligator clip contacts a bare metal surface NOTE For this procedure refer to FIGURE 2 5 1 Perform the steps in Accessing XPO2 on page 9 2 Remove the inlet filter with tubing attached from the holder 3 Disconnect the battery wiring harness ...

Страница 13: ... the XPO2 On to ensure proper operation 3 Turn XPO2 Off 4 While holding the unit together position the XPO2 so that the back panel is facing up Detail A in FIGURE 2 3 5 Install four mounting screws and secure the two sections of the XPO2 together Detail A in FIGURE 2 3 Securely tighten 6 Remove the glue residue by applying 91 isopropyl alcohol and wiping with a clean cloth Make sure the surface is...

Страница 14: ... now in auto pulse mode NOTE The unit will remain in auto pulse mode until it is set to a lower flow setting or is turned off 4 Connect a standard hand held oxygen analyzer to the outlet port 5 Follow the directions provided by the analyzer manufacturer Some analyzers will also measure flow and pressure but only if connected to a continuous flow of oxygen so these options are not usable on this pu...

Страница 15: ...repair centers To access the Sub Alarm codes the unit first has to be in an alarm condition with multiple flow level LEDs illuminated While in this condition the Plus and Minus flow setting buttons should be pushed and held As long as both these buttons are being pushed a different set of flow LEDs will illuminate This alternate set of illuminated LEDs is the sub alarm code Once the Plus and Minus...

Страница 16: ...Code 7008 3 4 5 4 Code 7009 3 4 5 4 1 Code 700a 3 4 5 4 2 Code 700b 3 4 5 4 2 1 Code 700c 3 4 5 4 3 Code 700d 3 4 5 4 3 1 Code 700e 3 4 5 4 3 2 Replacement of the internal battery may eliminate this problem Code 700f 3 4 5 4 3 2 1 Replacement of the internal battery may eliminate this problem Code 7010 3 4 5 5 Replacement of the internal battery may eliminate this problem Code 7011 3 4 5 5 1 Code ...

Страница 17: ...NOTES Part No 1150748 17 XPO2 NOTES ...

Страница 18: ...NOTES XPO2 18 Part No 1150748 NOTES ...

Страница 19: ...LIMITED WARRANTY Part No 1150748 19 XPO2 LIMITED WARRANTY For warranty information please refer to the original owner s manual which came with this product or contact Invacare for more information ...

Страница 20: ...sissauga Ontario L4Z 4G4 Canada 800 668 5324 All rights reserved Trademarks are identified by and All trademarks are owned by or licensed to Invacare Corporation unless otherwise noted Torx is a registered trademark of Textron Inc Dawn is a trademark of The Proctor and Gamble Company 2008 Invacare Corporation Part No 1150748 Rev A 4 08 ...