Invacare® TDX® SP2 Series
SYMPTOM
PROBABLE CAUSE
SOLUTIONS
Battery indicator
flashes the charge
level is low - too
soon after being
recharged.
Batteries not
charged.
Weak batteries.
Have charger
checked.
Replace batteries
if necessary.
Refer to
Replacing
Batteries and/or
Battery Cables,
page 53
Motor “chatters”
or runs irregular.
Motor/gearbox
malfunction.
Stop use of
Wheelchair.
Contact
Provider/Invacare
for Service.
Joystick erratic or
does not respond
as desired.
Damaged motor
coupling.
Electrical
malfunction.
Controller
programmed
improperly.
Contact
Provider/Invacare
for Service.
Contact
Provider/Invacare
for Service.
Contact
Provider/Invacare
to have controller
reprogrammed.
SYMPTOM
PROBABLE CAUSE
SOLUTIONS
Wheelchair does
not respond to
commands.
Electrical
malfunction.
Contact
Provider/Invacare
for Service.
Power indicator
off - even after
recharging.
Poor battery
terminal
connection.
Have terminals
cleaned.
11.2.1 Battery and Charger Tips
The following “Do’s” and “Don’ts” are provided for your
convenience and safety.
DON’T
DO
Don’t perform any
installation or maintenance
without first reading this
manual.
Read and understand
this manual and any
service information that
accompanies a battery and
charger before operating
the wheelchair.
Don’t perform installation
or maintenance of batteries
in an area that could be
damaged by battery spills.
Move the wheelchair to a
work area before cleaning
terminals, or opening
battery box.
Don’t make it a habit to
discharge batteries to the
lowest level.
Recharge as frequently as
possible to maintain a high
charge level and extend
battery life.
74
60101877-B
Содержание TDX SP2 series
Страница 86: ...Notes...
Страница 87: ...Notes...