Invacare® TDX® SP
7.5
Specifications
WEIGHT LIMIT
MODEL
MOTOR
Adult
Junior
TDX SP HD
Up to 400 lbs
(181 kg)
N/A
TDX SP
Up to 300 lbs
(136 kg)
Up to 150 lbs
(68 kg)
TDX SR
Up to 300 lbs
(136 kg)
Up to 150 lbs
(68 kg)
TDX SR HD
4 Pole HD
Up to 400 lbs
(181 kg)
N/A
7.6
Positioning the Wheelchair in the Vehicle
This wheelchair MUST be in a forward facing position during travel
in a motor vehicle.
The recommended clear zones for wheelchair seated occupants
restrained by both pelvic and upper torso belt(s) and only by a pelvic
belt are shown in the diagrams and described below.
Frontal Clear Zones (FCZ) need to be LARGER when upper torso
belt(s) are NOT used.
The rear clear zone of 16 inches (41 cm) is measured from the
rearmost point on an occupant’s head.
The frontal clear zone is measured from the frontmost point on an
occupant’s head and is 26 inches (66 cm) with pelvic and upper-torso
belt(s) and 37 inches (94 cm) with ONLY a pelvic belt.
The frontal clear zone may not be achievable for wheelchair-seated
drivers.
The estimated seated height (HHT) from the ground or floor to
the top of the wheelchair-seated occupant’s head ranges from
approximately 47 inches (119 cm) for a small adult female to about
61 inches (155 cm) for a tall adult male
7.7
Securement Points
ITEM
DESCRIPTION
A
TIE DOWN BRACKETS
52
1143190-N-03