SECTION 9—BATTERIES
70
5. Remove the 1/4-20-inch hex flange nut that secures the bracket of the POSITIVE
battery cable to the POSITIVE (+) battery post of the battery.
6. Remove the 1/4-20-inch hex flange nut and 1/4-20 x .88 hex flange screw that secures
the NEGATIVE battery cable to the NEGATIVE(-) battery post of the battery.
7. Discard the existing battery.
8. Position battery connector bracket or wiring harness onto the new 22NF battery as
shown.
9. Secure the NEGATIVE battery cable to the NEGATIVE (-) battery post with existing
1/4-20 x .88-inch hex flange screw and 1/4-20 hex flange nut. Securely tighten.
10. Secure the bracket of the POSITIVE battery cable to the POSITIVE (+) battery post with
existing mounting screw and 1/4-20-inch hex flange nut. Securely tighten.
11. Position each battery terminal cover over top of each battery terminal.
12. Secure battery terminal covers in place with one tie-wrap.
13. Install batteries into wheelchair. Refer to Removing/Installing the Batteries From/Into
FIGURE 9.1
Bracket of POSITIVE
Battery Cable
Battery Connector
Bracket
POSITIVE (+)
Battery
Terminal/Post
1/4-20 x .88-inch
Hex Flange Screw
22NF Battery
NEGATIVE (-)
Battery Cable
NEGATIVE (-)
Battery
Terminal/Post
1/4-20-inch Hex
Flange Nut
Battery Wiring Harness
FOR FRONT
BATTERY
FOR REAR
BATTERY
Bracket of POSITIVE
Battery Cable
NEGATIVE (-)
Battery Cable
BLACK Battery
Terminal Cover
RED Battery
Terminal Cover
Tie-Wrap
BATTERY COVER AND
TIE-WRAP DETAIL
BLACK
Cover
RED
Cover
Tie-Wrap