Invacare®Scanbeta NG™
Check the electronic wiring for the motors (wires not squeezed).
Check that the enclosing of the motors are intact (no cracks that enables fluids
to pierce).
Check the sections of the mattress support by using the hand control to activate
all functions of the moving parts.
Check the function of the brakes.
Check that the bed end attachments are properly mounted (steady) and that
the bed ends are securely attached.
Check the locking function(s) of the side rails.
6.5
Dismounting the motor
IMPORTANT!
–
If the bed is standing up on its castors, a minimum of
three persons are needed to dismount the motor for
height adjustment. Two need to hold up the mattress
support while one is dismounting the motor.
–
As an option, the bed can be put on the side. The
motor can then be dismounted by only one person.
IMPORTANT!
Before dismounting the motor.
–
Disconnect the bed from the mains socket.
–
Disconnect all motor cables from the electronics.
•
The lock cam over the connectors, on the electronics
box, must be removed before contacts can be released.
•
Certain models also have bolts on the motors, which
need to be loosened.
1.
Press the snap locks (one in each end of the lock cam) with a
screwdriver and remove the cam.
2.
Pull out the motor plug from the electronics.
Dismount the motor
1.
Lift the brace up, and fold it away from the cotter pin.
2.
Pull out the cotter pin.
3.
Repeat the step in the other end of the motor.
4.
Remove the motor.
IMPORTANT!
–
Keep a firm grip and be careful not to drop the motor.
Mount the motor
1.
Place the motor in its insertion place and push the cotter pin
through all holes.
2.
Lift the brace up and fold it over the cotter pin to lock. See
illustration in section “Dismount the motor”.
After mounting the motor
1.
Put the motor plugs back in the electronics. Each socket
is marked with a function. Make sure every function is
connected to the right motor. If you are not certain
which function the motor has, the article No gives more
information.
2.
Mount the lock cam and make sure it is secured with
the snap locks.
3.
Check that all motors and the hand control have correct
functionality.
6.6
Dismounting the mattress support
IMPORTANT!
Before dismounting/mounting
–
Make sure the brakes are activated.
–
At least two persons are needed to dismount/mount
the mattress support. Both ends of the mattress
support have to be supported.
45
0
A
Dismounting the mattress support
1.
Disconnect the cables between the electronics and the motors
to the shear arms.
2.
Release the latches by pulling out the red buttons
A
and turning
them 45° clockwise.
3.
In the foot end; lift the mattress support, releasing it from the
buckles.
4.
In the head end: pull out the shear arms gliding blocks from the
guideways on the mattress support.
5.
Lift off the mattress support from the shear arms.
Mounting the mattress support
1.
Lift the mattress support in over the shear arms.
2.
Lift the mattress slightly in the foot end and gently slide the shear
arms gliding blocks into the guideways on the mattress suport
(head end).
3.
Gently lower the foot end, fitting the buckles on the mattress
support on the ends of the shear arms crossbar and press until
you hear a clicking sound.
4.
Plug in the cables from the motors to the electronics.
16
1570628-B
Содержание Scanbeta NG