Relative humidity
20 % to 90 %, not condensing
Atmospheric pressure
860 hPa to 1060 hPa
11.6 Electrical system — Models equipped with electric tilt and backrest
Voltage supply
U in 100–240 Voltage, AC, 50/60 Hz (AC = Alternating current), DC 24 V (DC = Direct current)
Maximum current input
I in max. 5 Ampere (battery = max 1,3 Ampere)
Intermittent (periodic
motor operation)
10 % 2 min ON / 18 min OFF
Protection class
CLASS II equipment
Applied Part complying with the specified requirements for protection against electrical shock
according to IEC60601-1.
Degree of protection
The control unit, external power supply, motors and hand controls are protected according to
IPx6. See label and label on each electric device for correct IP class. The lowest IP-classification
decides the overall classification of the system.
IPx6
- The system is protected against water projected from any direction (not high pressure).
For full details contact your Invacare authorized provider.
11.7 Electromagnetic compliance (EMC)
General information
Products with electronic equipment needs to be installed and used according to the EMC information in this user manual.
WARNING!
Portable and mobile communications equipment can affect the operation of this product.
This product must not move unintentionally while being submitted to Electromagnetic interference
– If this should happen; the product must immediately be taken out of use and checked by a qualified technician.
This product has been tested and complies with the EMC limits specified by IEC 60601-1-2 for Class B equipment.
This product has a very low emission and should not interfere with other equipment.
However, if other devices nearby should react inexplicably, run and stop this product and observe the devices.
•
If nothing happens with the other devices, then this product is not causing the error.
•
If other devices are behaving in an inexplicable manner, then this product is causing the error. Solve the problem
by moving or increasing the distance between them.
11.8 EMC - Manufacturer´s declaration
Test
Basic Standard
Compliance Level / Frequency Range
Mains PortConducted Emissions
EN 55011
Group 1 Class B
Radiated Emissions
EN 55011
Group 1 Class B
Mains Harmonics
EN 61000-3-2 +A1 + A2
Class A
Voltage Fluctuations
EN 61000-3-3
dc%/dmax/d(t)/Pst
Electrostratic Discharge
EN 61000-4-2
±2/4/6kV Contact, ±2/4/8kV Air
Radiated Field Immunity
EN 61000-4-3
3V/m, 80% 1kHz AM, 80MHz - 2.5 GHz
EFT/Bursts
EN 61000-4-4
±2kV L/N/E/LNE
Surge Immunity
EN 61000-4-5
±0.5/1kV L-N, ±0.5/1/2kV L-E/N-E
Conducted RF Immunity
EN 61000-4-6
3V, 80% 1kHz AM, 0.15MHz - 80MHz
1488727-R
57