SECTION 13—CONTROL PANEL
51
FIGURE 13.1
Replacing Control Panel
DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical
outlet before servicing.
1. Remove the control panel. Refer to Removing Control Panel on page 50.
2. Remove the P.C. board. Refer to Replacing P.C. Board on page 54.
3. Remove the transformer. Refer to Replacing Transformer Assembly on page 56.
4. Remove the On/Off switch. Refer to Replacing On/Off Switch on page 57.
5. Remove the flow meter. Refer to Replacing Flowmeter on page 58.
6. Remove the hour meter. Refer to Replacing Hour Meter on page 59.
7. Remove horn and palnuts that hold the barbed outlet fitting.
8. Discard the existing control panel.
9. Install new control panel by reversing STEPS 1-7.
10. Run unit and inspect for leaks. Refer to Leak Test on page 71.
11. If no leaks detected, turn power Off (O) and unplug the unit.
12. Reinstall cabinet. Refer to Removing Cabinet on page 25
Circuit Breaker (STEP 11)
Hour Meter (STEP 11)
1/8” I.D. Pressure
Sensor Tubing (STEP 5)
Oxygen Sensor Tube
(SensO
2
Models Only)
Nine or Ten
Pin
Connector
(STEP 8)
Upper Mounting
Screws
ON/OFF
Switch (I/O)
(STEP 11)
Patient Outlet
Barbed fitting
Top and Bottom
Barbed fittings
8-Pin Connector
(STEP 9)
Lower Mounting
Screws
Control Panel
Transformer
Harness
Connector
(STEP 10)