6.4 Cleaning the Cabinet Filter
CAUTION!
Risk of Damage
To avoid damage to the internal components of the unit:
-DO NOT operate the concentrator without the filter installed or with a dirty filter.
CAUTION!
Risk of Damage
To avoid breaking the filter assembly:
-DO NOT remove the filter material from the filter holder.
There are two cabinet filters, one located on each side of the cabinet.
There are two ways to clean the cabinet filter:
l
Vacuum the filter.
l
Remove the filter door assembly and wash the filter.
6.4.1 Vacuum the Filter
1. Use the vacuum hose to remove the dirt from the filter.
2. If the dirt is not removed with the vacuum, follow the instructions in the next procedure for
washing the filter.
6.4.2 Washing the Filter
Fig. 6-1 Washing the Filter
1. Press the filter assembly on both sides of the concentrator and slide up to remove to clean as needed.
Environmental conditions that may require more frequent inspection and cleaning of the filter
include, but are not limited to: high dust, air pollutants, etc.
2. Clean the cabinet filter assembly with a vacuum cleaner or wash with a mild liquid dish detergent (such as
Dawn) and water. Rinse thoroughly. Thoroughly dry the filter assembly .
3. Inspect for fraying, crumbling, tears, and holes. Replace filter assembly if any damage is found.
4. Reinstall the cabinet filter assembly on both sides of the concentrator.