Invacare IH6300 Скачать руководство пользователя страница 1

Owner’s Operation and Maintenance Manual

DEALER:

 This manual MUST be given to 

the user of the tub.

USER:

 BEFORE using this tub, read this 

manual and save for future reference.

For more information regarding 

Invacare products,  parts, and services, 

please visit www.invacare-ccg.com

Supine Bath Tub

IH6300 Fixed Height

IH6300ADH Adjustable Height

AFTER INSTALLATION, PLEASE CALL CUSTOMER SERVICE AT

1-800-668-2337 TO ARRANGE FOR AN IN-SERVICE ON THE TUB

Содержание IH6300

Страница 1: ...ub read this manual and save for future reference For more information regarding Invacare products parts and services please visit www invacare ccg com Supine Bath Tub IH6300 Fixed Height IH6300ADH Ad...

Страница 2: ...ERSONNEL BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY OCCUR THE INITIAL SET UP OF THIS TUB MUST BE PERFORMED BY A QUALIFIED TECHNICIAN PROCEDURES OTHER THAN THOSE DESCRIBED I...

Страница 3: ...ISINFECTING THE TUB 12 Disinfecting the Tub Using the Closed Loop System 12 SECTION 3 OPERATION 13 Overview 13 Preparing the Tub for Use 13 Operating the Tub 14 Using Water Temperature Setting 14 Oper...

Страница 4: ...re Each Use 25 Every 80 150 Baths 25 Every Six Months 26 Maintenance for IH6300ADH Adjustable Height ONLY Glide Pads 26 Removing Soap Film and Hard Water Deposits 26 Removing Calcium Deposits Scale an...

Страница 5: ...equipment If RFI causes erratic behavior PUSH the Red Power Switch OFF IMMEDIATELY DO NOT turn the Power Switch ON while transmission is in progress MAINTENANCE Maintenance MUST be performed ONLY by...

Страница 6: ...nfectant BEFORE use Refer to Owner s Manual for proper disinfecting procedures 1095523 Rev A 1 02 00 WARNING For disinfecting use ONLY DO NOT use for showering Refer to Owner s Manual for proper disin...

Страница 7: ...s 1245 mm 49 inches 1245 mm Weight Tub shipping weight 725 lbs 329 kg 725 lbs 329 kg Tub empty 700 lbs 318 kg 700 lbs 318 kg Tub w 103 gal of water max water capacity 1575 lbs 714 kg 1575 lbs 714 kg F...

Страница 8: ...er Disinfectant Wand Disinfectant Cabinet Water Temperature Thermometer Shower ON OFF Tub Fill Spout Disinfectant On Off Whirlpool ON OFF Model IH6300 Fixed Height Tub Tub Controls UP Down Whirlpool A...

Страница 9: ...performed by a qualified electrician and is in compliance with local electrical codes Electrical Grounding Instructions Model IH6300 Supine Tub MUST be direct wired to wall using hot neutral and grou...

Страница 10: ...ield protection when disinfecting or cleaning your tub ALWAYS handle disinfectants cleaning compounds and powdered sanitizers with extreme care Physical or mechanical damage caused by improper handlin...

Страница 11: ...pool jets Severe pump damage will result DO NOT operate this unit without the guard over the water intake suction fitting The guard is a safety device that reduces the potential hazard of hair or body...

Страница 12: ...2 Depending on the type of whirlpool jets that are in your tub close all jets by performing one of the following Turn the center of the jet clockwise to a full stop Push the center of the jet in 3 Cl...

Страница 13: ...d before preparing the tub for use NOTE This procedure must be performed after disinfecting the tub to finalize the disinfecting process 1 Open the whirlpool inlet by pushing the inlet plug in the cen...

Страница 14: ...g Water Temperature Setting Flow Water Thermometer Measures the temperature of water coming from the tub fill spout or shower massage wand Water Temperature Control Valve Sets and maintains constant t...

Страница 15: ...esired level NOTE If hydromassage use is anticipated water level MUST be completely above whirlpool jets NOTE Before starting hydromassage open the whirlpool inlet and whirlpool jets Using Hydromassag...

Страница 16: ...e than three cycles up and down within a ten minute period Damage to the actuator may result NOTE The supine tub stops automatically when maximum height adjustment Up or Down is reached Raising the Su...

Страница 17: ...lling tub 2 Carefully remove the tub from carton 3 Remove tub from shipping pallet by removing screws from leg mounting pads 4 Model IH6300 ONLY Remove the screws that secure the rear access panel to...

Страница 18: ...mounting pads with the holes drilled in STEP 4 WARNING DO NOT adjust height of legs more than 1 inch 25 4 mm leg will disassemble from leg mounting pad and serious injury may occur 6 Level the tub by...

Страница 19: ...ed electrician and is in compliance with local electrical codes Before installing and connecting the tub make certain that the electric power source is SHUT OFF to avoid electrical shock NOTE For this...

Страница 20: ...aker Only IH6300ADH Adjustable Height Use 15 Amp Hospital Grade GFCI Receptacle Only DETAIL B ELECTRICAL HOOK UP DETAILS DETAIL A ELECTRICAL HOOK UP LOCATION Wall Tub Wall Pump Motor BLACK WHITE GREEN...

Страница 21: ...the supply and installation of a Ground Fault Circuit Interrupter GFCI Before installing and connecting the tub make certain that the electrical power source is shut off to avoid electrical shock NOTE...

Страница 22: ...r to make the elbow flush with the floor For Model IH6300ADH this will allow smooth up and down movement of the hose Otherwise drain tubing and fittings could be crushed at low point of operation 1 In...

Страница 23: ...wn to clear lines of debris prior to connection Valves and spray heads may become inoperable and or damaged if lines are not properly cleared Items damaged in this way are not covered under the warran...

Страница 24: ...ve Tub Fill Valve Disinfectant Line Flow meter Disinfectant Valve Jet Pump In House Water Service RADA Mixing Valve Optional Pressure Regulator Back Flow Preventer Disinfectant Injector Tub Fill Check...

Страница 25: ...Information NOTE For parts information please contact Invacare Continuing Care Group Technical Service at 866 616 2631 NOTE Everysixmonthsorasnecessary haveaqualifiedtechnicianperformathoroughinspect...

Страница 26: ...H6300ADH Adjustable Height ONLY Glide Pads Removing Soap Film and Hard Water Deposits Use a commercially available cleaner such as Invacare s Heavy Duty Cleaner to thoroughly spray the exposed surface...

Страница 27: ...blue The use of these types of cleaners will cause the tub to stain Never use cleaners containing bleach Use of these types of cleaners will dry out the rubber seals and gaskets and the tub will not f...

Страница 28: ...could result in injury to the tub operator or resident ALWAYS wear rubber gloves an apron and face shield protection during disinfecting cleaning and rinsing procedures In case of eye or skin contact...

Страница 29: ...ry 18 Note the number of ounces in the graduated plastic bottle 19 Let the bucket pail fill to one gallon 20 Turn the disinfectant valve to the off position 21 Remove the disinfectant draw tube from t...

Страница 30: ...SECTION 5 MAINTENANCE Supine Bath Tub 30 Part No 1096926 Maintenance Record DATE MAINTENANCE PERSON TYPE OF MAINTENANCE PERFORMED NOTES...

Страница 31: ...art No electrical power to tub Check and reset breaker or GFCI System not prepared Be sure water level is above jets Check that inlet plug on tub is depressed Switch hose disconnected Reconnect hose t...

Страница 32: ...L WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS LABOR EXPENDABLE ITEMS I E SHOWER HOSES VALVES O RINGS SEALS PARTS BROKEN DUE TO ABUSE OR IMPROPER USE ITEMS THAT BECOME INOPERABLE DUE TO LACK...

Отзывы: