Chapter 5 : Diagnostics
9
A thermal circuit breaker must be installed between the batteries and the
Power Module - as close as possible to the batteries - to protect both the
batteries and the system wiring. If the two batteries are permanently wired
together (single battery box), the best position for the circuit breaker is
between the two batteries. If the batteries are separated (individual battery
boxes), each battery requires a circuit breaker.
A 30-40 Amp slow-acting, thermal type circuit breaker is suggested.
The thermal circuit breaker should have a trip rating no higher than the
current limit of the Power Module. Check thoroughly to ensure that it
provides the necessary degree of motor protection.
Содержание Hydra C650PW
Страница 29: ...29 ...
Страница 31: ...Installation Manual SHARK DK REMA Series of Control Units 2 Button Remote by GBK80260 Issue 1 June 2004 ...
Страница 34: ...7 7 Electromagnetic Compatibility EMC 48 7 8 Contact Details 49 ...
Страница 36: ...GBK80260 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 48: ...GBK80260 Issue 1 1 07 2004 14 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 64: ...GBK80260 Issue 1 1 07 2004 30 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 66: ...GBK80260 Issue 1 1 07 2004 32 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 71: ...Chapter 7 Appendices 37 Dynamic SHARK Looms For DK REMA DK PMA Power Module ...
Страница 74: ...GBK80260 Issue 1 1 07 2004 40 Dynamic SHARK Connector Kits and Adapters DK PMA Power Module Shown ...
Страница 84: ...Installation Manual DK PMA SHARK Power Module by GBK80262 Issue 1 June 2004 ...
Страница 87: ......
Страница 89: ...GBK80262 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 91: ...GBK80262 Issue 1 1 07 2004 4 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 113: ...GBK80262 Issue 1 1 07 2004 26 Dynamic SHARK DK PMA Looms ...
Страница 116: ...Chapter 7 Appendices 29 Dynamic SHARK Connector Kits and Adapters ...