GBK80260 : Issue 1 – 1/07/2004
28
5.2
Diagnostics Tools
While SHARK indicates the abnormal condition, a hand held programmer or
the PC-based SHARK Support Tool will provide more detailed information on
the fault.
Hand Held Programmer
Plugging a hand held programmer into the SHARK Control Unit when an
abnormal condition exists will cause the fault to be displayed. A 4-digit code
will be displayed which indicates the condition. The first two digits provide the
flash code number. The second two digits provide more specific diagnostics
information that is suitable for repair technicians.
In some cases, viewing a history of any abnormal conditions that occurred
previously on the system may be useful in diagnosing the current condition.
This can be done by entering the Fault Log from the
Diagnostics
menu.
Usage statistics are also available from this menu.
DYNAMIC
Wizard
Wizard is the preferred diagnostics tool in the workshop environment,
providing a full fault history and verbal descriptions of each flash and
associated servicing code.
If after analyzing the data, the condition cannot be diagnosed, it is possible to
print or save a Status Report for further analysis or distribution to a service
center.
Содержание Hydra C650PW
Страница 29: ...29 ...
Страница 31: ...Installation Manual SHARK DK REMA Series of Control Units 2 Button Remote by GBK80260 Issue 1 June 2004 ...
Страница 34: ...7 7 Electromagnetic Compatibility EMC 48 7 8 Contact Details 49 ...
Страница 36: ...GBK80260 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 48: ...GBK80260 Issue 1 1 07 2004 14 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 64: ...GBK80260 Issue 1 1 07 2004 30 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 66: ...GBK80260 Issue 1 1 07 2004 32 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 71: ...Chapter 7 Appendices 37 Dynamic SHARK Looms For DK REMA DK PMA Power Module ...
Страница 74: ...GBK80260 Issue 1 1 07 2004 40 Dynamic SHARK Connector Kits and Adapters DK PMA Power Module Shown ...
Страница 84: ...Installation Manual DK PMA SHARK Power Module by GBK80262 Issue 1 June 2004 ...
Страница 87: ......
Страница 89: ...GBK80262 Issue 1 1 07 2004 2 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 91: ...GBK80262 Issue 1 1 07 2004 4 THIS PAGE IS LEFT BLANK INTENTIONALLY ...
Страница 113: ...GBK80262 Issue 1 1 07 2004 26 Dynamic SHARK DK PMA Looms ...
Страница 116: ...Chapter 7 Appendices 29 Dynamic SHARK Connector Kits and Adapters ...