Invacare® Formula™ CG Seating
WARNING!
–
If the telescoping tubes need to be extended greater
than two inches, then the seat must be repositioned
rearward to ensure stability - otherwise injury and/or
damage to the wheelchair and surrounding property
may result.
7.10.1
Adjusting Telescoping Front Rigging Support
1.
Remove the two cap screws
A
, spacers
B
and threaded blocks
C
securing the telescoping front tube
D
to the side rail
E
.
2.
Slide existing telescoping front rigging support to one of six
depth positions.
3.
Use the existing two cap screws, spacers and threaded blocks to
secure the telescoping front tube to the side rail.
4.
Repeat steps 1 to 3 on the opposite side, if desired.
7.10.2
Replacing Telescoping Front Rigging Support
1.
Remove the two cap screws
A
, spacers
B
and threaded blocks
C
securing the telescoping front tube
D
to the side rail
E
.
2.
Remove existing telescoping front rigging.
3.
Perform one of the following:
•
Reclining Seating Systems Only - Refer to Detail “A” and
the list below to determine the mounting position for the
new telescoping front tubes.
– 16 inch and 19 inch Seat Depth
F
– 17 inch and 20 inch Seat Depth
G
– 18 inch and 21 inch Seat Depth
H
– 22 inch Seat Depth
I
–
•
Non-reclining Seating Systems - The telescoping front tubes
can be positioned using one of six mounting positions.
Proceed to step 4.
4.
Use the existing two cap screws, spacers and threaded blocks to
secure the telescoping front tube to the side rail
5.
Repeat steps 1 to 4 on the opposite side, if desired.
7.11
Installing Adjustable Angle Flip-up
Footplate Hinge
1.
Position footplate hinge
A
on the footrest support tube
B
at
the desired height.
2.
Position mounting screw
C
, washers
D
, spacer
E
, and locknut
F
on the footrest support as shown.
3.
Flip the footplate hinge to the up position.
The footplate hinge will fall to the down position.
56
1143155-R-01
Содержание Formula CG Elevate