Check points
Possible issue
Action
Comments
OK
The hand grips are
damaged.
•
Replace the hand grips.
Before installing the new
hand grips, clean the
tubes with an oil-free
grease solvent such as
window cleaning fluid.
The hand grips can
easily be rotated.
•
Replace the hand grips.
Before installing the new
hand grips, clean the
tubes with an oil-free
grease solvent such as
window cleaning fluid.
The adjustment knob
does not lock properly.
•
Replace the knobs and the
locking sleeves.
Check that hand grips
and handle tube are
fixed properly and are
free from defects.
The handle tubes are
bent or damaged.
•
Replace the handle bars.
The frame is damaged
or defect.
•
Scrap the device.
Check that the frame is
free from defects.
The coating is damaged
or scratched.
•
Use touch-up colors to repair
damages.
Touch-up colors are
available as spare parts.
The backrest or the
backrest attachments
are damaged.
•
Replace the backrest.
The screws on the
backrest or the backrest
attachments are loose.
•
Tighten the screws.
Check that seat and
backrest are fixed
properly and are free
from defects.
The seat is damaged.
•
Replace the seat.
2.4 Cleaning
The product can be cleaned and disinfected using commercially available detergents.
IMPORTANT
– Cleaning must be done at least once a year.
– Do not use an abrasive detergent cleaner.
– Do not use a steam cleaner.
– Never use acids, alkalies or solvents.
•
The product can be washed in a washing machine or by hand.
•
The washing and drying temperature may not be higher than 60º C.
•
High-pressure cleaner or ordinary jet of water may be used. However, do not aim the jet towards bearings, draining
holes or labels.
•
Clean with damp cloth and mild detergents.
•
Clean the wheels with a brush (no wire brush!).
•
After washing, wipe the product dry.
Disinfection
Disinfect the product by wiping down all generally accessible surfaces with disinfectant.
1649877-A
7