Invacare® Carroll CS Series™
8 Troubleshooting
8.1 Troubleshooting
Effect
Possible cause
Verification
Corrective action
Floor lock not engaged.
Green plunger on floor lock
post is visible.
Activate floor lock - green
plunger should NOT be visible.
Insufficient static friction
between floor lock foot
bumper and floor.
Floor surface is slippery
and/or foreign membrane
between bumper and floor.
Clean the surrounding floor
and remove loose material
under the bed.
Bed does not stay in place
Floor lock not functioning.
Floor lock is stuck in one
position.
Contact Invacare Continuing
Care Service Department
1-800-668-2337.
Bed is not level
Hi/Lo actuators are unevenly
extended
Bed deck is not parallel with
floor
Raise bed to highest position
or
Lower bed to lowest position
Actuator not working.
System Fault
Needs Reset
Flashing LEDs on ACP, Bed
function will not operate. The
control box may beep when
pendant function is pressed.
To Reset - hold the pendant
bed up and bed down buttons
for 5+ seconds. The bed will
beep and stop beeping when
reset is complete. Then press
the bed up button. If the
bed beeps, the actuators are
out of sync. Continue rising
to the highest position to
synchronize the actuators.