9 Troubleshooting
9.1 Identifying and repairing faults
WARNING!
– Only personnel having received the necessary instruction or training by Invacare must perform service and
maintenance on the product.
Symptoms
Faults
Solution
Mast/Base joint loose.
Refer to Assembling the Mast to the
Base in the Setup section.
Patient lift feels loose.
Tie - Rods are loose.
Contact your Dealer or Invacare
representative.
Casters/Brakes noisy or stiff.
Fluff or debris in bearings.
Clean casters from fluff and debris.
Noisy or dry sound from pivots.
Needs lubrication.
Refer to Lubricating the Lift in the
Maintenance section.
Electric actuator fails to lift.
Hand-control or actuator connector
loose.
Connect hand control or actuator
connector. Ensure connectors are seated
properly and fully connected.
Battery low.
Charge batteries. Refer to Charging the
battery in the Usage section.
RED emergency stop button pressed IN.
Rotate RED emergency stop button
CLOCKWISE until it pops out.
Battery not connected properly to
control box.
Reconnect the battery to the control
box. Refer to Charging the battery in the
Usage section.