S Y S T E M O P E R A T I O N
CONFIGURE SYSTEM Screen Detailed
3-30
ATS-515, -525 & ATS-615 & -625 Interface & Applications Manual
Section J:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CONFIGURE SYSTEM
CONFIGURE SYSTEM Screen
Detailed
CONFIGURE
SYSTEM Screen
Descriptions
This screen displays the system’s software version and allow configuration of system settings.
Use the Rotary Encoder to scroll through the menu selections. Press the Rotary Encoder to
activate a selection.
Part
Function
ESC
program button
Press to exit the
CONFIGURE SYSTEM
Screen
and return to the
TOP MENU
Screen
without saving changes.
DONE
program button
Press to
SAVE
configured settings and return to the
TOP MENU
Screen
.
IEEE 488 ADDRESS
(0 - 30) [GPIB]
Set IEEE General Purpose Interface Bus address: range is 0 to 30
(default is 9).
SERIAL BAUD RATE
(600 - 38.4k)
Accepted baud are 600; 1200; 2400; 4800; 9600; 19,200; (default
is) 38,400.
ST/EOT/SFF Polarity:
(0=neg 1=pos)
Set signals sent to/from System: to positive or negative (neg. is
default). If set to positive, then the user must supply LOW signal to
the energized pin.
Auto start MANUAL
OPERATION
Sets what screen the System displays, at boot, after
System Startup
Screen
:
If On:
MANUAL OPERATION
Screen
displays on startup.
If Off:
TOP MENU
Screen
displays on startup.