MANUAL TEST SETUP Screen
3-14
ATS-515, -525 & ATS-615 & -625 Interface & Applications Manual
MANUAL TEST SETUP Screen
MANUAL TEST
SETUP Screen
Detailed
MANUAL TEST
SETUP Screen
Descriptions
This screen allows for 12 Manual Test Setups that cycles through three setpoints (HOT,
AMBIENT, and COLD).
Each Manual Test Setup can be saved as part of a Test Setup File (numbered “1” through
“12”).
Each Test Setup file consists of:
1. a Manual Test Setup (detailed here)
2. a Program Test Setup (for more detail, see
3. a Define Parameters setup that is common to both the Manual and Program Test Setup (for
more detail, see
.
Part
Function
ESC
program button
Press to exit the
MANUAL TEST SETUP
Screen
and return to
the
MANUAL OPERATION
Screen
.
SELECT
program button
Press to cycle through the TEST SETUPS “0” - “12”.
ATTENTION
- Parameter changes and Test Setups should be
saved under Test Setup “1” through “12”.
- DO NOT save under Test Setup “0” as “0” is the
System default. Test Setup “0” will be over written
because it always reflects the last Test Setup used.
- The desired Test Setup should be selected
BEFORE parameter changes are made.