Intersurgical PMH7000 Скачать руководство пользователя страница 10

Instruction For Use                                                           PMH7000                                                         

       

Version 2

   

 
 

Revision 03                                                                  23 July 2019    

 

                                                   Page 10 of 20 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

d

 in meters 

(m) can be estimated using the equation applicable to the frequency transmitter, where

 P

 is the maximum output 

power rating of the transmitter in watts (W) according to the transmitter manufacturer. 

 
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies. 
 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

reflection from structures, objects and people. 

 

 

3.5 Compliant Accessories 

Dual temperature probe (REF 7700000) and single heated wire adaptor lead (REF 5600000) or dual heated wire adaptor 
lead (5601000). 

 

Patient systems, chambers, adaptor leads and dual temperature codes commercially available from other manufacturers 
have not been fully validated. As the PMH7000 Humidifier base incorporates a shunt fuse, the PMH7000 is not compatible 
for use with electrical adaptors containing a shunt fuse.

 

 

3.6 Settings for Ventilated Patients  

IPPV mode:

  

When the gas flow is in the range of 10L/min to 60L/min, an absolute humidity of at least 33mg/L is delivered.  

NPPV mode:  

When the gas flow is in the range of 10L/min to 60L/min, an absolute humidity of at least 10mg/L is delivered.

 

 

 

3.7 Capacity of Humidifying Chamber 

Effective capacity of humidifying chamber is less than 500mL. 
 

4.  Temperature  

The airway temperature and chamber temperature are servo controlled by the Humidifier base.   
 

4.1 Set Temperature Range 

Patient Side 

30

°

C to 40

°

(increments 1

°

C) 

Chamber Side 

26

°

C to 43

°

(increments 1

°

C) 

Temperature setting of patient side vs. chamber side temperature is 

in relationship within the range of -4

°

C to +3

°

C. 

 

4.2 Temperature Display 

Patient side temperature or chamber side temperature is displayed by switching using         , which is 
displayed in Chapter 5.1 while operated in manual mode. 

Patient Side Temperature Display Range 

10

°

C to 47

°

C increments 0.1

°

C  

Chamber Side Temperature Display Range 

10

°

C to 45

°

C increments 0.1

°

C  

 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 

Содержание PMH7000

Страница 1: ...03 23 July 2019 Page 1 of 20 Intersurgical Humidifier Heater Base PMH7000 Instructions for Use Distributed by Intersurgical Ltd Crane House Molly Millars Lane Wokingham Berkshire RG41 2RZ UNITED KING...

Страница 2: ...late Temperature 7 3 1 8 Heater Wire Resistance 7 3 2 Physical Dimensions 7 3 2 1 Size Without Humidification Chamber Installed 7 3 2 2 Weight 7 3 3 Environmental 7 3 3 1 Operating Temperature 7 3 3 2...

Страница 3: ...Left Side View 14 6 2 Left Side Panel Description 14 7 ASSEMBLY WITH BREATHING CIRCUIT 15 8 SET UP AND OPERATION OF THE HUMIDIFIER BASE 16 9 MAINTENANCE REPAIR 17 9 1 Basic Checks 17 9 2 Dual Tempera...

Страница 4: ...vice 2 1 Warnings Cautions and Notes In this manual a WARNING describes the actions necessary to prevent injury or loss of life A CAUTION describes the actions to be taken to protect the equipment fro...

Страница 5: ...n cause patient injury if not acted upon WARNING ALARMS No alarms sound or alarm location lights on the front panel can be activated while the system is shut down or while the protective thermostat is...

Страница 6: ...ing Radiation Power On Off IPPV Mode NPPV Mode Alarm Silencer Button Do not discard 3 Specifications WARNING EXPLOSIVE HAZARD Do not use this equipment in the presence of flammable gases or anaestheti...

Страница 7: ...10 to 95 Non condensing That includes the transportation time 3 3 5Atmospheric Pressure 700 to 1060 hPa 3 4 Electromagnetic Interference PMH7000 complies with limitations as specified in IEC 60601 1 2...

Страница 8: ...y network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacturer s declarat...

Страница 9: ...ordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess electromagnetic environment due to fixed RF...

Страница 10: ...0 Humidifier base incorporates a shunt fuse the PMH7000 is not compatible for use with electrical adaptors containing a shunt fuse 3 6 Settings for Ventilated Patients IPPV mode When the gas flow is i...

Страница 11: ...est and alarm sound test as well When it is turned off alarm sound verification will be activated only once in short time Note In order to verify the functionality of the alarm system during the power...

Страница 12: ...nk quickly in green and display patient temperature setting At this time temperature setting changes can be made by pressing up and down button Press the button once and the temperature value changes...

Страница 13: ...rmittently LED Dual temperature probe Flashes red The dual temperature probe 7700000 is disconnected or faulty in this case a high priority audible alarm sounds intermittently LED Heater plate Flashes...

Страница 14: ...Connector Temperature probe connecting port 3 Inspiratory Expiratory Heater Wire Connector Heated wire adaptor connecting port NOTE The power cord cannot be detached from the heater base The power co...

Страница 15: ...g port Patient side temperature sensor Connect to the patient side temperature sensor connecting port on the breathing circuit Heater wire connection Inspiratory side Connect the connector of inspirat...

Страница 16: ...connect the cloverleaf connector to inspiratory heated wire limb and oval connector to the expiratory heated limb When using a single heated wire systems connect the electrical adaptor 5600000 to the...

Страница 17: ...n the dual temperature probe is exchanged or replaced be sure to turn off the power of the Humidifier base After exchange or replacement re activate the Humidifier base 12 Safe Process to stop operati...

Страница 18: ...abnormal conditions during operation Each alarm shows the abnormal conditions by illuminating an LED on the front panel and sounding the alarm Alarm sound can be muted by the mute button for 2 minute...

Страница 19: ...e inter lock is activated It is locked when alarm state continues for 10 minutes Turn off the device and restart it The device is not powered on No 1 Device fused 2 Heater base fault 3 Overheat protec...

Страница 20: ...Instruction For Use PMH7000 Version 2 Revision 03 23 July 2019 Page 20 of 20...

Отзывы: