7000 Series Respiratory Humidifier 

1

Power Switch

2

Temperature Display

3

Mode Change

4

Mode Display LED’s

5

Scroll Buttons

6

Set Temperature Button

7

Status Indicator LED’s

8

Alarm Silence Button

MODE

SET

!

°

C

MANUAL

AUTO

ALM

T

emp 

C

onnection

1

2

3

4

7

5

6

8

Turn on Humidifier

 by briefly pressing power switch (1) Unit will default to last mode 

used.

Select desired mode

 by holding down Mode Button (3) repeat until desired mode is 

selected.

IPPV Mode: 

This mode is preset for intubated patients. The temperature for the 

patient side is controlled to 40°C and chamber side to 37°C.
 

NPPV Mode: 

This mode is preset for patients using a NIV (Non Invasive Ven-

tilation) interface. The temperature for the patient side is controlled to 34°C and 

chamber side to 31°C.

MANUAL mode: 

This mode allows the selection of temperatures for patient side and 

chamber side. 
The set values can be changed by following the instructions in Section 5.3.4 of the Instruc-

tions For Use.

Temperature Display (2). 

The display shows either chamber outlet temperature or 

patient temperature (whichever is selected as default).  To display the alternate measured 

temperature, hold down the alarm silence button (8) for 2 seconds, the display will then 

revert after 10 seconds. 

Alarms.

 An audible alarm will sound. A status indicator LED (7) on the front panel will 

highlight the location of the fault. The audible alarm can be muted for 2 minutes by pressing 

the alarm silence button (8). 
If an Alarm state is left unresolved for 10 minutes, the humidifier will enter stand-by, with the 

display showing “000”. Switch humidifier off and on to exit stand-by.