To conserve battery life, the meter will automatically shut off after 5 minutes unless plugged
into a power source or while conducting a graph measurement.
To recharge the battery, plug the mini USB cable into the slot on the top of the meter. Plug
the USB A end of the cable into a PC, USB wall plug, USB power bank or any standard
mobile phone USB A charging device. The ILT710 will always turn on when power is supplied
via the mini USB. It will take approximately 4 hours to charge when the battery is drained.
The meter can be used while charging the battery.
Battery replacement cannot be done in the field as it requires soldering to the board which
can cause damage by untrained personnel. Please contact ILT for battery replacement.
6. SENSOR
The Meter and sensor have been paired and calibrated to create a NIST traceable calibrated
light measurement system. Swapping sensors between meters can cause inaccurate results.
To attach the sensor, locate the female BNC connector on the top side of the meter. Align the
connectors, press down and twist to lock the sensor in place. The meter is designed so that
the sensor can be detached when not in use, or left attached at all times.
The sensor is not waterproof and should not be submerged. Liquids should be used very
sparingly when cleaning the sensor to prevent possible leakage.
Sensor replacements are available, however, replacing a sensor requires a new lux/fc
calibration. Consult ILT before swapping or replacing sensors.
7. USB PORT
On the top right side of the meter is the Mini USB port. The USB port is used for both
charging the device and for downloading stored data to the PC. ILT recommends use with the
supplied USB cable or any good quality mini USB cable.
8. CALIBRATION
ILT typically recommends an annual calibration for all equipment, however, the end user may,
through their own QC process, create their own calibration cycle.
Service: Before returning any equipment to ILT for service, please use the ILT website to
obtain your Return Material Authorization number (RMA):
Website: https://www.intl-lighttech.com/form/rma-form
Phone: 978 818 6180 x 118
ILT710 Flicker Meter