background image

Wireless Light-Level Monitoring

ILT1000

Error Handling Condition

Result

Wireless access point goes out of 
range

When the monitoring computer was moved out of the 
wireless network range, while actively accessing the 
ILT1000 devices, the HMPS2U required power-cycling to 
properly re-identify the ILT1000 devices.  This monitoring
computer did not experience any errors or reboot 
requirements, but HCC needed to be run after power 
cycling the HMPS2U to reconnect the devices.  Under 
normal conditions, based on accessing the HMPS2U via 
the existing wireless network, this condition should not 
occur normally.  Nonetheless it is recommended that 
system design ensures wireless network range is 
maintained.

Monitoring station/computer returns 
from power down mode.

HCC finds the devices, but requires manual “Connect” 
action on each device before connection can be 
reestablished.

Reboot of the monitoring 
station/computer.

HCC finds the devices, but requires manual “Connect” 
action on each device before connection can be 
reestablished.

Summary:

-

Pros

o

Works as advertised, providing robust wireless-to-USB bridging and supporting the 

ILT1000.

o

Simple installation and setup, after making initial connection to the router and not a 

laptop as depicted in the instructions.

o

Fairly good speed (multiple readings per second when interfacing to 3 ILT1000’s) 

through the wireless interface.

-

Cons

o

Devices, other than printers, cannot be auto-connected upon reboot or return from 

low-power modes.

Wired-to-USB Bridge Configuration

For all configurations below, testing was performed on a Windows 8.1 system.

The Wired-to-USB Bridge configuration testing was performed as illustrated in Figure 2 above 

and reiterated below in Figure 10 (showing the TP-LINK TL-WR702N as the wireless router).  In 

this configuration, the monitoring computer is talking to a new or existing wireless network, 

which has a wired connection to a USB bridge. 

12

Содержание ILT1000

Страница 1: ...Wireless Light Level Monitoring ILT1000 ILT1000 Application Note Wireless Light Level Monitoring 1...

Страница 2: ...is focused on wireless monitoring the same principles apply to wired Ethernet monitoring Disclaimer The application note below describes ILT s experience integrating 3rd party devices with its ILT100...

Страница 3: ......

Страница 4: ......

Страница 5: ...llation 1 The installation calls for a wired Ethernet connection for the initial configuration but an Ethernet cable is not provided in the package 2 A reboot is required after installation 3 The test...

Страница 6: ...Wireless Light Level Monitoring ILT1000 6...

Страница 7: ...t does not automatically connect to them Note that HCC also displays printer icons as opposed to something more generic for the ILT1000 s To connect to an ILT1000 the particular device is selected as...

Страница 8: ...Level Monitoring ILT1000 Figure 5 Device connected After manually connecting the remaining two devices all show as connected Figure 6 and all show up within Windows Device Manager as USB to Serial dev...

Страница 9: ......

Страница 10: ......

Страница 11: ......

Страница 12: ...reestablished Reboot of the monitoring station computer HCC finds the devices but requires manual Connect action on each device before connection can be reestablished Summary Pros o Works as advertise...

Страница 13: ......

Страница 14: ......

Страница 15: ......

Страница 16: ......

Страница 17: ...to ensure that wireless connectivity is maintained while accessing the NUS device Monitoring station computer returns from power down mode With Connect Device Automatically is selected See Figure 15...

Страница 18: ......

Страница 19: ...tically connected upon start up or reboot Summary Pros o Works as advertised with the combination of the wireless router and Ethernet USB Bridge providing robust wireless to USB bridging and supportin...

Отзывы: